第1章 生体信号の基礎 各種生体信号

バイオメディカル情報工学
第二部 生体情報技術入門
各種生体信号
Biosignal
Range
心弾動図
Ballistocardiogram
(BCG)
0-7 mg 0-40
Accelerometer, strain gage
0-100
μm
0-40
Displacement (LVDT, Linear Variable
Differential Transformer )
第2部 生体情報技術入門
膀胱圧
Bladder pressure
0-10
Strain gage manometer
第1章 生体信号の基礎
1-100
cm
H2O
血流
Blood flow
1-300
ml/s
0-20
Flowmeter (electromagnetic or
ultrasonic)
(血圧値と心拍数の計測)
生体情報学講座
陳 文西
Biosignal
Sensor
直接
10-400 0-50
mm Hg
Strain-gage manometer
間接
25-400 0-60
mm Hg
Cuff, auscultation
静脈血圧
Blood pressure,
venous
0-50
0-50
mm Hg
Strain gage
動脈血圧
Blood
pressure,
arterial
Biosignal
Range
Freq., Hz
Sensor
皮質脳波(ECoG)
Electrocorticogram
10-5000 μV
0-150
Brain-surface or depth
electrodes
Range
Freq., Hz
PO2
30-100
mm Hg
0-2
筋電図(EMG)
Electromyogram
0.1-5 mV
0-10,000
Needle electrodes
PCO2
40-100
mm Hg
0-2
眼電図(EOG)
Electrooculogram
50-3500 μV
0-50
Contact electrodes
1-3 mm
Hg
0-2
網膜電図(ERG)
Electroretinogram
0-900 μV
0-50
Contact electrodes
0.10.4mm
Hg
0-2
電気皮膚反応(GSR)
Galvanic skin
response
1-500 kΩ
0.01-1
Skin electrodes
血液pH
Blood pH
6.8-7.8
pH units
0-2
胃筋電図(EGG)
Electrogastrogram
10-1000 μV
0-1
Skin-surface electrodes
心拍出量
Cardiac output
4-25
liter/min
0-20
Dye dilution, Fick
0.5-80 mV
0-1
Stomach-surface electrodes
0.01-250
Skin electrodes
0-100 cm
H2O
Strain-gage
0.5-4 mV
Gastrointestinal
pressure胃腸圧
0-10
心電図(ECG)
Electrocardiogram
1-50 g
0-1
Displacement system, LVDT
脳波(EEG)
Electroencephalogram
5-300 μV
0-150
Scalp electrodes
Gastrointestinal
forces胃腸力
胃pH
Gastric pH
3 - 13 pH
units
0-1
pH electrode, antimony
electrode
血液ガス
Blood gases PN2
PCO
第一章 生体信号の基礎
Sensor
Freq.,
Hz
Specific electrode, volumetric
or manometric
Specific electrode
1
バイオメディカル情報工学
第二部 生体情報技術入門
Biosignal
Range
Freq.,
Hz
Sensor
神経電位
Nerve potentials
0.01- 3 mV
010,000
Surface or needle electrodes
心音図
Phonocardiogram (PCG)
5-2000
Dynamic
range 80 dB,
threshold
about 100
μPa
Microphone
容積脈波 Plethysmogram
(volume change)
Varies with
organ
0-30
Displacement chamber or
impedance change
Flow rate
0-600
liter/min
0-40
Pneumotachograph head
and differential pressure
Respiratory
rate
2-50
0.1-10
breaths/min
Strain gage on chest,
impedance, nasal thermistor
Tidal
volume
50-1000
ml/breath
0.1-10
Above methods
32-40 ℃
90-104 ℉
0-0.1
Thermistor, thermocouple
呼吸機能
Respiratory
functions
体温
Body temperature
心電図と動脈血圧
ECGとABP(Arterial Blood Pressure)
第一章 生体信号の基礎
心電図(ECG)
Electrocardiogram
心電図と呼吸
ECGとRESP(respiration)
2
バイオメディカル情報工学
自発脳波 Spontaneous EEG
(Electroencephalogram)
第二部 生体情報技術入門
誘発脳波
Evoked, Event-related EEG
聴覚誘発電位 (AEP)
Auditory Evoked Potential
視覚誘発電位(VEP)
Visual Evoked Potential
体性感覚誘発電位(SEP)
Somatosensory Evoked Potential
筋電図(EMG)
Electromyogram
体温(BT)
Body Temperature
睡眠時体温の変動
第2部 生体情報技術入門
第一章 生体信号の基礎
第1章 生体信号と計測技術
3
バイオメディカル情報工学
第二部 生体情報技術入門
血圧(BP),心拍数(HR),体重(BW),体温(BT)
収縮期圧 SBP (Systolic)
拡張期圧 DBP (Diastolic)
計測手法
第一章 生体信号の基礎
信号の多様性
計測システムの基本構成
4
バイオメディカル情報工学
第二部 生体情報技術入門
信号とノイズ
生体計測の特殊性
• 組織・器官破壊や生理状態の乱れを最小限にする
• 生理量の変化は短時間の成分にも長時間の成分
にも意味があり、計測システムの広い周波数応答
性と高い生物的・化学的安定性が要求される
• 組織や器官が柔らかく壊れやすく、異物に対して拒
絶反応のため、生体と直接接触するセンサ素材に
生物親和性が要求される
• 人間に意思と感情があり、強制や拘束、不快や痛
みを最小限にする
• Signal
– the component of a variable that contains
information about the object quantity
• Noise
– a component unrelated to the object quantity
• Signal ↔ Noise
– not defined by physical nature but by the intention of
the observer
• Signal-to-Noise Ratio (SNR)
SNR dB  10 log10
– P and A indicate power and RMS amplitude,
respectively
ECGと各種雑音
Power line interference
Baseline drift
PS
A
 20 log10 S
PN
AN
ノイズの種類
• Thermal Noise
– Random thermal agitation relevant to temperature
– Uniform distribution of power density
• 1/f Noise
EMG interference
Motion artifact
– Many natural phenomena
– Power density is inversely
frequency
proportional
to
the
• Interference
– Electromagnetic coupling - power line, fluorescent
lamps
• Artifact
– Superimposed on the object quantity and caused by
external factors such as motion – skin-electrode
contact
第一章 生体信号の基礎
5
バイオメディカル情報工学
第二部 生体情報技術入門
電磁ノイズ(EMI)
ElectroMagnetic Interference
1/f ノイズ
ノイズの確率分布
amplitude
amplitude
uniform
distribution
time
amplitude
amplitude
amplitude
probability
probability
density
function
生体内圧力の計測
血圧BP
Gaussian
distribution
time
amplitude
第一章 生体信号の基礎
probability
probability of
noise occurs in
such amplitude
6
バイオメディカル情報工学
第二部 生体情報技術入門
生体内の圧力
臓器系
圧力名
循環器
心室内圧
心臓の収縮と拡張
(大)動脈圧
左心室の収拡によって(大)動脈内の圧力、血圧BP
呼吸器
泌尿器
消化器
その他
生体内圧の変動範囲
説明
静脈圧
静脈血管内の圧力。右心房近傍の中心静脈圧CVP
肺動脈圧
右心室の収拡によって肺動脈内の圧力PCWP
微小血管内圧
直径20-250umの血管内圧、微小循環動態
気道内圧
呼吸に伴って生じる気道内の圧力、呼吸機能
胸腔内圧
呼吸に伴って生じる胸腔内の圧力、呼吸機能
消化管内圧
胃あるいは腸管の内圧
膀胱内圧
膀胱内に貯留する尿液の圧力
尿道内圧
排尿するための圧力
頭蓋内圧
頭蓋骨内部の圧力ICP。脳圧、脳髄液圧とも言う
眼圧
眼球内の圧力
子宮内圧 第1章 生体信号と計測技術
羊膜内の羊水圧
第2部 生体情報技術入門
圧力
•
•
•
•
•
•
•
•
•
単位面積当たりに働く力、単位パスカル(Pa)
1Pa=1N/m2
水銀柱ミリメートル(mmHg)
水柱センチメートル(cmH2O)
1mmHg=133.322Pa
1cmH2O=98.066Pa
絶対圧=真空を基準とする=大気圧+ゲージ圧
ゲージ圧=大気圧(101,325Pa)を基準とする
生体内圧力=一般的にゲージ圧(gauge pressure)
第一章 生体信号の基礎
血圧
• 心臓血管循環系には体循環と
肺循環がある、静脈と動脈
• 心臓が血液を送り出す際に血
管の壁にかかる圧力
• 一般的に動脈にかかる圧力
• 通常は上腕の動脈で血圧を測
定(心臓の位置と同じレベル)
7
バイオメディカル情報工学
心臓血管循環系の血圧
第二部 生体情報技術入門
血圧計測法の分類
• 観血法、侵襲法(invasive)、直接法
– 動脈カテーテル法(arterial catheter)
• 非観血法、非侵襲法(noninvasive)、間接法
– 間欠法(intermittent method)
• コロトコフ(聴診)法(Korotkoff method)
• オシロメトリック(容積振動)法(Oscillometric method)
• 超音波ドップラー法(ultrasound Doppler method)
– 連続法(continuous method)
• トノメトリ法(tonometry method)
• 容積補償法(unloading method)
• 脈波伝播時間法(pulse wave transit time method)
動脈カテーテル法
半導体ゲージセンサ
第2部 生体情報技術入門
第一章 生体信号の基礎
第2章 生体信号計測技術
32/56
8
バイオメディカル情報工学
第二部 生体情報技術入門
コロトコフ法
光ファイバセンサ
第2部 生体情報技術入門
カフ(内圧)振動法
第2部 生体情報技術入門
第一章 生体信号の基礎
第2章 生体信号計測技術
第2章 生体信号計測技術
34/56
血管の静的圧と容積の関係
35/56
9
バイオメディカル情報工学
指式の容積振動法
トノメトリ法による連続血圧計測
第一章 生体信号の基礎
第二部 生体情報技術入門
超音波ドップラー法
容積補償法ー概念図
10
バイオメディカル情報工学
容積補償法ー記録例
第二部 生体情報技術入門
脈波伝播概念図
脈波伝播時間と血圧の関係
演習課題
血圧値と心拍数の計測
第一章 生体信号の基礎
11
バイオメディカル情報工学
第二部 生体情報技術入門
内容・課題
• 内容
– 使用設備:
• K音式手動血圧計1種類+手首式と上腕式自動血圧計各2種類
– 実験プロトコル:
• 「安静時→運動中→回復中」の順で各機種各3回計測
• 運動方法は体力に応じて個人別選択(現地足踏み、マルチバイク、階段上下)
• 血圧値(SBPとDBP)と心拍数(HR)の計測結果を実験用紙に記録
– チーム構成:
• 4グループ分け、4~5人/グループ
• 被験者=1人/グループ、記録者=1人/グループ、実験者=その他
• 課題
–
–
–
–
横方向:三回計測結果の平均値と標準偏差を求める
縦方向:計測部位毎の計測結果の平均値と標準偏差を求める
オンライン情報サイトを利用し、実験に使った血圧計の仕様を調べる
計測部位、身体状態、血圧計種類によって生じた計測値(BPとHR)
の差について、考えられる可能な原因を考察する
第一章 生体信号の基礎
12