Download - Tnmsc.com

TENDER FOR SUPPLY AND INSTALLATION OF SUCTION APPARATUS AC/DC &
MANUAL, BAINS CIRCUIT (PEDIATRIC), MULTICHANNEL MONITOR AND
MONITOR PULSE OXIMETER WITH PLETHYSOMOGRAPHY EQUIPMENT TO TNEMRI
TENDER NO. 5407/TNMSC/ENGG/2014, dt.10.06.2014
RESULTS OF EVALUATION AND FINALLY SHORTLISTED BIDDERS.
SUCTION APPARATUS AC/DC & MANUAL
Sl.
No
Bidders
Quoted
1.
M/s.
Model/
Make
Vive Eterno – II
Solutions
Responsive bidders
Reason for rejection
Technical
evaluation
report
Responsive
Commercia
l evaluation
report
Responsive
Non
Responsive
Non
Responsive
-------
AC/DC
Pedal
Supreme
make
2.
M/s. SISCO
Latex
EUROVAC
AC/DC
As per technical evaluation report:
1. Weight is more than 5 Kgs.
Private
2. AC& DC indication not available.
Limited 3. Battery backup is only 1.30 against
the tender requirement of 4 hrs.
4. Wall mountable option not available.
As per commercial evaluation report:
The bidder did not furnish any certificate
such as end user certificate. Annual
turnover statement and cash credit limit,
despite given chance .
MULTICHANNEL MONITOR
Sl.
No
Bidders
Quoted
Model/Make
1.
M/s.
Trans G3G, General
Health
Care Meditech
India Pvt. Ltd., Inc.,
P.R.China
Responsive bidders
Commercial
Technical
evaluation
evaluation
report
report
Responsive
Responsive
Reason for rejection
--------
Sl.
No
Bidders
Quoted
2.
M/s.
Model/Make
M
M Vismo
Medical
2703,
Systems
Kohden
Responsive bidders
Commercial
Technical
evaluation
evaluation
report
report
PVM NonNikon responsive
Responsive
Reason for rejection
As per the demo report, the model
quoted is not meeting the following
points:
1. Battery backup is only for 3 hrs
against the tender requirement
of 4 hrs.
2. Spo2 range is only 50 to 100%
instead on 0 to 100%.
3. Power consumption is 100 watt
against the tender requirement
of less than 50 watts.
4. The model is not simple to use,
operate and maintain.
3.
M/s.
Indian PM1200JET,
Medical
Systems
Riransa
Nonresponsive
Nonresponsive
&
Not demonstrated and as per the
commercial evaluation report, the
bidder did not furnish end user
certificate for their manufacturer
Technologies
despite given chance.
4.
Sign NonBPL responsive
As per the demo report, the model
M/s.
Smart
Solokrafts
S10,
Industries
Medical
points:
Technologies
1. I/O Communication – COMI –
Responsive
quoted is not meeting the following
RS 232 is not available.
Pvt. Ltd.
2. System is not able to upgrade
to EtCo2 & printer in future.
3.
The equipment can operate
only in 0 to 40 deg C and
humidity of 90% against the
tender requirement of 10 to 45
deg C and humidity of 100%.
Sl.
No
Bidders
Quoted
Model/Make
Responsive bidders
Commercial
Technical
evaluation
evaluation
report
report
Reason for rejection
4. The equipment is sensitive to
fluctuation.
5.
M/s
Schiller Truscope II.
Healthcare
BioLight
Nonresponsive
Responsive
As per the demo report, the model
quoted is having weight of 5.610
kgs against the tender requirement
India Pvt. Ltd.,
of weight of less than 4.5 Kgs.
6.
M/s.
Vin YK-8000C,
Xuzhou
Biotech
Yongkang
Systems Ltd., Electronic
Science
Technologies
Co. Ltd.
Responsive
Responsive
--------
MONITOR PULSE OXIMETER WITH PLETHYSOMOGRAPHY
Sl.
No
1.
Bidders Quoted
M/s.
Health
Trans G1B+, General
M/s.
Responsive bidders
Commercial
Technical
evaluation
evaluation
report
report
Responsive
Responsive
Reason for rejection
--------
Care Meditech Inc.,
India Pvt. Ltd.,
2.
Model/Make
P.R.China
Indian Masimo SAT
Responsive
Responsive
OXM NonM/s
Schiller Argus
responsive
Healthcare India Plus, BioLight
Pvt. Ltd., Responsive
--------
Medical Systems 901+, Masimo
& Technologies
3.
As
per
the
wrapping
available.
Hence
not
by
the
specialists.
Bains Circuit (Pediatric) – No response report,
protective
recommended
demo
is
not