Meters • UV Minder - Solar Light Company

Meters • UV Minder
Model 3D
Radiometer with UVA
and UVB Sensors
Applications
•
•
•
•
•
•
SPF testing
Cosmetics Research
Dermatology
Environmental Monitoring
Environmental Research
Agriculture and Forestry
Features and Benefits
•
•
•
•
•
•
•
•
High Sensitivity
Battery Operated
LCD Display
Optional Six foot Shielded Extension Cable for Analog Output
NIST Traceable Calibration
Radiometric Units
HOLD Toggle, Freezes the Present Reading
AUTO SHUT OFF Meter Turns Off After 8 Minutes with Near
Zero Reading
• Rugged Carrying Case
• BAT Low Battery Indicator
The UV Minder® Model 3D UV intensity meter is capable of measuring the
UVA and UVB spectra. The UV meter shows the intensity of UVB (also called
SUV – Sunburning UV) in Minimal Erythemal Doses per Hour (MED/Hr), the
accepted clinical measure for sunburn potential.
The UVA detector measures UV intensity in the range between 320 and 400
nm, displaying UVA irradiance in milliwatts per square centimeter.
Also available as the UV Minder® Model 3D-XP Xeroderma Erythemal UV
Intensity Meter for measuring very low UV intensity to 3 decimal places. Please
specify wavelength range required when ordering. The UV meter is very easy
to use with its sensors mounted on the meter, making UV measurement easy,
and its light weight makes it very easy to handle. The UV Minder UV meter is
a very versatile instrument.
The UVA and SUV sensors are conveniently mounted on the top of the meter,
making it easy to point the sensor at the target and register the result by
selecting the desired sensor. A green flashing indicator light on the keypad
shows which sensor is active. Press the HOLD button to freeze the current
reading. To release, press the HOLD button again. Turning off the meter also
resets the hold status. The HOLD function is useful if the detector needs
to be in a place where the display can not be read. Once the detector is in
place press the HOLD button. You can now move the detector and meter to a
position where you can read the display.
An analog output is provided at the bottom of the meter. This allows the
user to continuously monitor and record the intensity irradiated on the active
detector. Recording can be accomplished by a strip chart recorder or an A/D
card connected to a computer. Plug the supplied 0.141 inch phone plug into
the analog output jack. The red lead is positive and the black lead is negative.
The calibration factor for the analog output is 0.1[Volt/(MED/HR)] or 0.1[Volt/
(mW/cm2)].
Meters • UCV Minder
Model 3D
Specifications
Dectector Inputs Two
UVB (SUV) 0...19.99 MED/Hr
Input Ranges
UVA
0...19.99 mW/cm2
Resolution 0.01 MED/Hr or mW/cm2
Operating Environment 32 to 120 °F (0 to +50 °C) No Precipitation
Power Source 9 volt Alkaline Batteries
Power Consumption <300µA
Battery Life >40 Hours Continuous Use
External Power 9-12V AC or DC Adapter
Program Control 4-Button Keypad
Size (WXDXH) 3.75”x5.5”x3.4” (9.5x14x8.6cm)
Fig. 1. Model 3D Spectral Response
LCD Size 3.5 Digit
Weight 8 oz. (227 grams)
Ordering Information
UV Minder® Model 3D
Xeroderma Erythemal UV Intensity Meter
UV Minder® Model 3D-XP for Measuring Very Low UV Intensity to 3
Decimal Places. Please Specify.
References
1
2
The biological effects of UV-A radiation - Edited by F. Urbach and R.W. Gange, Praeger Publishers,
New York, 1986
Nichodemus F., “Self study manual on optical radiation measurements”, NBS Technical Note 910-1
(1976).
SL/Meters_Model 3D_10/2014
100 East Glenside Avenue • Glenside, PA 19038 • USA • P 1.215.517.8700 • F 1.215.517.8747 • www.solarlight.com • [email protected]