Analytic performance of the ABX PENTRA DX 120 compared to the manual microscope method, the ADVIA 2120 and the SYSMEX XE 2100 in the cell counting of body fluids P. Danise (1), B. Talento (2), A. Rovetti (1), C. Ricciardi (1), C. Esposito (2), A. Di Palma (1), D. Avino (1), E. Tournier (3), M. Pastore (3), M.G. Pirofalo (1) (1) Haematology Laboratory, Umberto I Hospital, Nocera Inferiore (SA), Italy (2) Clinical Pathology Laboratory, Umberto I Hospital, Nocera Inferiore (SA), Italy – (3) Horiba Medical, Montpellier, France Introduction Results Body fluids cell count is performed both in routine and emergency situations. Often single laboratories lack dedicated specialists and therefore BF analysis may be conducted by staff who do not have equal or necessary skill level to do this diagnosis. Manual counting, that constitutes the basis of the reference method, has an elevated analytic variability, by reason of the features of the method itself and the limited number of counted cells. Such a method, used in non standardized conditions and by personnel who perform a few numbers of tests per year, may show reduced accuracy and reproducibility. In these conditions the use of an automated counting system can be advantageous in term of accuracy and reproducibility, as is the norm when an automated procedure replaces a manual one. Furtheremore we evaluated the possibility of counting cells distinct from polymorphonuclear and mononuclear cells (neoplastic and mesothelial cells) as a screening method for the subsequent correct microscope identification. SYSMEX XE 2100 Manual count 4-5074 53-5207 1-67 5-6770 77-5199 1-42 5-5878 108-6122 1-76 10-5000 80-5000 1-60 Tab. 2 – Correlations of the total cell counts ABX Pentra DX120 Advia 2120 SYSMEX XE 2100 SIEMENS Advia 2120 SYSMEX XE 2100 Manual count 0.984 0.992 0.990 0.994 0.989 0.994 Ascitic F. Pleuric F. Cerebrospinal F. MN ABX Pentra DX120 MN Manual PMN ABX Pentra DX120 PMN Manual OTHERS ABX Pentra DX120 OTHERS Manual 26-99 21-99 0-100 3-86 20-99 0-100 1-74 1-79 0-100 14-97 1-80 0-100 6-86 4-80 0-50 10-85 7-82 0-13 Tab. 4 – Correlations of polymorphonuclear (PMN), mononuclear (MN), and other (LIC) cells between ABX Pentra DX 120 and manual count Ascitic F. Pleuric F. Cerebrospinal F. Material & Methods We examined 100 Body Fluid samples: 15 cerebrospinal (CSF), 42 pleural and 43 ascitic fluids, coming from routine and emergency services at the Hospital Central Laboratory. Total counting of each sample has been conducted with all systems. Mononuclear, Polymorphonuclear and Large Immature Cells (LIC) (OTHERS) counting, representing cells different from the first two categories, has been performed only with the HORIBA Medical system and the manual method. Here the procedure specific to each apparatus: (HORIBA Medical) After cytochemical staining with chlorazol black, the cells are injected into a double hydrodynamic focused cytometer that aligns cells and allows cell by cell analysis by impedance (cell volume) and absorbance (cytoplasmic and nuclear structure), through the determination of the diffraction of a polychromatic light source following cell passage. (SYSMEX) The cells, after specific staining, are transported with a laminar flow into the detection chamber where they cross a monochromatic light beam in a hydrodynamic focused flow. Adapted optical systems (photodiodes and photomultipliers) detect the diffused light at different angles of each cell. The parameters measured by the optical bench in the Sysmex systems are Forward Scattered Light and Side Scattered Light. ADVIA 2120 SIEMENS Advia 2120 (SIEMENS) The total nucleate cell (TNC) counting in Ascitic and Pleural fluids is obtained in the basophil/lobularity channel, after lysing of RBC and the leukocyte cytoplasmic membrane, excluding basophils. For CSF specimen, previously sphericized and fixed cells, are to be differentiated and counted by mean of three optical signals (high-angle scatter, low-angle scatter and absorbance); the signals are then converted into digital form. MANUAL COUNTING For fresh cell counting we used the Burker’s chamber. For mononuclear, polymorphonuclear cells and LIC a slide was prepared using Aerospray Hematology 7150® cytospin to deposit the cells. After May Grumwald-Giemsa staining we performed microscope analysis. MN 0.848 0.902 0.522 (P=0.06) PMN 0.844 0.898 0.543 (P=0.05) OTHERS 0.827 0.881 0.242 (P=0.38) Conclusions The total cell counting comparison between different analyzers showed a good correlation of data without any relevant difference among the evaluated technologies. Equally optimal results were obtained comparing the 3 instruments with the manual method. As expected, the cell number ranges in the diverse body fluids showed values definitely lower in CSF. This fact is a known cause of reduced accuracy within the performances of the analyzers when faced with this kind of specimen. Indeed the concentrations of cells are extremely lower in the CSF than in the blood for which these technologies are meant. The evaluation of polymorphonuclear, mononuclear, and other cells performed in the comparison of the ABX Pentra DX 120 (HORIBA Medical) and the manual method showed that: (1) The correlation values, obtained for the pleural and ascitic fluids (Tab. 4), make possible to consider this analyser for counting in routine use. This is even truer if the results are correlated to the variance of the microscopic counting performed by non specialized staff. (2) The CSF correlation is less acceptable, due to the essentially very low concentration of cells. Even in this case it is difficult to evaluate, without further specific studies, what could be the variance of manual counting in real life. (3) The counting evaluation of cells different from polymorphonuclear and mononuclear showed surprisingly positive results. The CSF was an exception for the reason mentioned above. Such good performances appear even more noteworthy if we consider the exclusive screening role of the automated count that expects the manual microscopic methodologies to produce the correct identification of the counted cells. In conclusion automated body fluid counting has overcome the pioneer and experimental phase, and it is now offering homogeneous instrumental performances that can be a prelude to routine use. The CSF cell count in automation presents nowadays aspects that have to be improved, related to the low number of cells. The comparison between the ABX Pentra DX 120, the other technologies and the manual counting showed instrumental overlapping capabilities; moreover it highlighted the possibility to identify polymorphonuclear and mononuclear cells with satisfactory safety and to offer a reliable screening method for the evaluation of the presence of other cells in the examined fluids. REFERENCES 1.Marthe et al. “ Automated Flow Cytometric Analysis of Blood Cell in Cerebrospinal Fluid” Am J Clin Pathol 2004; 121: 690 – 700. 2.Harris et al. “ The ADVIA 2120 Hematology System: Flow Cytometry- Based Analysis of Blood and Body Fluid in the routine hematology laboratory” Laboratory Hematology 2005; 11:47-61. 3.Brown et al. “ Validation of body fluid analysis on the Coulter lh 750” Laboratory Hematology 2003; 9: 155 – 159. ISLH 2011 Here we present the comparison of a new counting mode proposed by HORIBA Medical utilising Siemens’ and Sysmex’s technologies and the manual method in order to evaluate the accuracy and the correlation between the methods. SYSMEX XE 2100 Ascitic F. Pleuric F. Cerebrospinal F. HORIBA Medical ABX Pentra DX120 Tab. 3 – Ranges (%) of polymorphonuclear (PMN), mononuclear (MN) and other (LIC) cells Aim ABX PENTRA DX 120 Tab. 1 – Ranges of the total cell counts performed
© Copyright 2024 ExpyDoc