DATE: 05/05/2014 DOC.MIE13175 TD 5000 APPLICATION GUIDE Rev 1.1.1 MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 2 of 54 REVISIONS N PAGE 1.0.5 All SUMMARY VISA DATE 18/01/2014 Issued Lodi 1.0.6 04/02/2014 Complete Revision Meneghin 05/05/2014 Added: Hot collar test Meneghin 1.1.1 All MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 3 of 54 SHORT FOREWORD ............................................................................................. 5 INTRODUCTION................................................................................................... 6 1 SAFETY AT WORK ........................................................................................... 10 1.1 INTRODUCTION TO SAFETY ................................................................................... 10 1. 2 SAFETY SYMBOLS .............................................................................................. 12 1.3 RISKY SITUATIONS .............................................................................................. 12 2 TEST SET DESCRIPTION AND USE .................................................................... 14 2.1 THE FRONT AND SIDE PANELS ............................................................................... 14 2.2 LOCATING TD 5000 .......................................................................................... 15 2.2.1 Mounting the TD 5000 module.............................................................. 15 2.2.2 Mounting the STS module ..................................................................... 16 2.2.3 Mounting the cable wheels ................................................................... 16 2.2.4 Preparing for testing: modules connections .......................................... 17 2.2.6 Preparing for testing: test cables connection ........................................ 20 2.2.7 Preparing for testing: connection to the power transformer ................ 23 2.3 POWER SUPPLY ................................................................................................. 25 2.4 POWER-ON ...................................................................................................... 25 2.5 GENERATOR CHARACTERISTICS ............................................................................. 26 3 TAN(Δ) AND CAPACITANCE MEASUREMENT BASICS ...................................... 30 3.1 CAPACITANCE AND TD MEASUREMENT OF A GENERIC TEST OBJECT .............................. 32 3.1.1 Tests execution ...................................................................................... 33 3.1.2 View Results........................................................................................... 34 3.1.3 Save results ............................................................................................ 36 3.1.4 Open results ........................................................................................... 36 3.1.5 Software settings ................................................................................... 36 4 TEST MODES .................................................................................................. 38 5 POWER TRANSFORMERS ................................................................................ 41 5.1 TWO WINDINGS TRANSFORMER ............................................................................ 41 5.1.1 Test connections .................................................................................... 42 5.1.1.1 Capacitance and TD measurement of CH, CHL or CH+CHL .......................... 42 5.1.1.2 Capacitance and TD measurement of CL, CHL or CL+CHL ............................ 42 5.2 THREE WINDINGS TRANSFORMER .......................................................................... 43 5.2.1 Test connections .................................................................................... 43 4.2.1.1 Capacitance and TD measurement of CH, CHL, CH+CHL, CHT, CH+CHT, CHL+CHT or CH+CHL+CHT........................................................................................ 43 5.2.1.2 Capacitance and TD measurement of CL, CHL, CL+CHL, CLT, CL+CLT, CHL+CLT or CL+CHL+CHT ......................................................................................... 44 MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 4 of 54 5.2.1.3 Capacitance and TD measurement of CT, CHT, CT+CHT, CLT, CT+CLT, CHT+CLT or CT+CHL+CHT ......................................................................................... 44 5.3 AUTOTRANSFORMER .......................................................................................... 45 5.3.1 Test connections .................................................................................... 45 5.3.1.1 Capacitance and TD measurement of CH .................................................... 45 5.4 AUTOTRANSFORMER WITH TERTIARY ..................................................................... 46 5.4.1 Test connections .................................................................................... 46 5.4.1.1 Capacitance and TD measurement of CH, CHT or CH+CHT.......................... 46 5.4.1.2 Capacitance and TD measurement of CT, CHT or CH+CHT .......................... 47 5.5 BUSHING ......................................................................................................... 48 5.5.1 Test connections, capacitance and TD measurement of C1 .................. 48 5.5.2 Test connections, capacitance and TD measurement of C2 .................. 49 5.5.3 Hot collar test ........................................................................................ 49 Disclaimer Every effort has been made to make this material complete, accurate, and up-to-date. In addition, changes are periodically added to the information herein; these changes will be incorporated into new editions of the publication. ISA S.R.L reserves the right to make improvements and/or changes in the product(s) and/or the program(s) described in this document without notice, and shall not be responsible for any damages, including but not limited to consequential damages, caused by reliance on the material presented, including but not limited to typographical errors. Copies, reprints or other reproductions of the content or of parts of this publication shall only be permitted with our prior written consent. All trademarks are the property of their respective holders. Copyright 2012© ISA S.R.L. Italy – All rights reserved. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 5 of 54 SHORT FOREWORD Dear TD 5000 user, I often wondered why user manuals are not very much used, even if they include valuable information. As me too I am a user of such manuals, the answer I have given myself is that valuable information are concealed somewhere in the thick thing, and I do not have time to waste to find it. So, either the manual is actually of help, or I ignore it. This is why I decided to separate the TD 5000 manual from the STS family manuals, so that all instructions related to this fundamental option of the STS family are kept together. Therefore, you will find here the following information: introduction to TD 5000 and the user guide. Have a good job with ISA test sets! Luca Biotti Q&A Manager NOTE: WINDOWS is a trademark of MICROSOFT inc. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 6 of 54 INTRODUCTION The products of the STS family allow performing all tests foreseen by international Standards on CTs, VTs, Power Transformers, and to measure the tan(δ), Power Factor and capacitance of any device with the TD 5000 option. The use of STS models is described in MIE12175 STS FAMILY – Application Guide. The STS family is made of three models: STS 5000, STS 4000 and STS 3000 light. All of them can operate with TD 5000; as there is no difference in the TD 5000 operation, in this document the STS family module will be called STS, without specifying the model. All test sets of the family are controlled locally, by keyboard, dedicated keys, control knob and display, or by a PC, with the STS_Pro program, which is a part of the TDMS software, which is provided with the test set. The STS module controls completely the TD 5000 option, which has no local selection. The capacitance and tangent delta measurement with TD 5000 can be performed on CTs, VTs, CBs, PTs and bushings. Tests are performed in accordance with the following IEC standards: EN 60044-1; EN 60044-2; EN 60044-5; EN 60044-7; EN 60044-8; EN 60076-1, and also in accordance with ANSI/IEEE C57.13.1. The instrument is housed in a transportable aluminium box, which is provided with removable cover and handles for ease of transportation. A foldable trolley is also available. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 7 of 54 The following picture shows TD 5000. On the test set are located three panels, with connectors. The following one is located to the right: the BOOSTER INPUT must be connected to the EXT. BOOSTER connector of the STS unit. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 8 of 54 The following panel is located to the left: the COMMUNICATION PORT is used to establish the communication with STS and to supply the electronic boards of the TD 5000 unit. The two measurement input IN A and IN B are used for the sensing of a current that can be involved or not in the capacitance or tan(δ) measurement, according with the test mode (GSTg-A, UST-B, and so on). In the following picture is shown the high voltage output MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 9 of 54 The block schematic of the STS with TD 5000 operation is shown in the following: Device Under Test IN A IN B METERS --------SWITCH MATRIX HV TRANSFORMER CPU GND TD 5000 BOOSTER CONNECTOR COMMUNICATION PORT CPU POWER AMPLIFIER DISPLAY STS GND During the operation, the test is selected on the STS LCD screen through the multi-function knob; then, STS sends instructions and power to TD 5000, which generates the high voltage. The HV return connection passes through the selection switches, which select the input A or B, and through the high accuracy measures, where the current is measured. Measurements are sent to STS, which computes the desired parameters, and displays them to the operator. Test results are kept in the STS memory, and can be transferred to a PC at a later time, along with settings. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 10 of 54 1 SAFETY AT WORK 1.1 INTRODUCTION TO SAFETY The Product TD 5000 hereafter described is an option of the STS family of products, including: STS 5000, STS 4000, and STS 3000. The following notes apply to TD 5000 with any of the above STS models. TD 5000 is manufactured and tested according to the specifications, and when used for normal applications and within the normal electrical and mechanical limits, it will not cause hazard to health and safety, provided that all standard engineering rules are observed, and that it is used by trained personnel only. The User should carefully read the instructions and the examples of this manual prior to operate the test set. This application manual is published by the Seller, to be used together with TD 5000, as described in the corresponding introductory manual. The Seller reserves the right to modify the guide without warning, for any reason. This includes also, but not only, the adoption of more advanced technological solutions and modified manufacturing procedures, and also the addition of other features, not available in the first release. The Seller declines any difficulties arising from unknown technical problems. The Seller declines also any responsibility in case of modification of TD 5000, or of any intervention not authorized by the Seller in writing. The warranty includes the repair time and the materials necessary to restore the complete efficiency of TD 5000; so, it does not include other burdens, such as the transport and customs fee. Under no circumstances the warrantee includes any cost that the User may have suffered because of TD 5000 unavailability and downtime. TD 5000 is CE marked, and has been tested to operate according to EN 61010-1, with the following operating conditions: . Pollution degree 2: normally, non-conductive pollution occurs; . Measurement category 2, for measurement inputs; . Altitude: less than 2000 m; . Operating temperature: - 10 to 55 °C; storage: -20 °C to 70 °C; MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 11 of 54 . Relative humidity: 5 to 95 %, without condensing; . Inputs/outputs protection: IP 2X: IEC 60529, for all but high voltage outputs; IP4X for high voltage outputs; . The test set is portable, using the handles. Would TD 5000 be used beyond these limits, the safety of the test set could be impaired, and the Seller would not be liable for any occurring problem. TD 5000 has been tested to match the EMI/RFI standards, as requested by the European Directive 2004/108/EC; Applicable Standard: EN61326 : 2006. However, the User should not carry a pacemaker. 1.2 Prior to testing TD 5000 generates a voltage that may be lethal to the unadvertised user. ATTENTION TD 5000 is used to test HV devices. During connections and disconnections, the device must be grounded: follow the safety procedures! When the test is in progress, DON’T TOUCH the terminals for any reason. In case of doubt, press the EMERGENCY pushbutton with mechanical lock which is located on the front face of STS! In order to avoid any danger, the device under test should have the following characteristics: . Connection cables must be those provided with the test set; . Connection points must be isolated and not accessible; . Input circuits must have an isolation degree at least equal to the one of the product. . The test area should be clean of loose metal items. . When performing HV tests, we recommend isolating the area, so that nobody could accidentally touch the test item. For added safety, we suggest to use the optional warning strobe light: it will keep on flashing until the test is over. Anyway, the test set can be programmed to generate a buzz. . When performing HV test, an additional precaution is to drive the START/STOP using the optional remote safety switch: with its 20 m long cable, you can stay apart from TD 5000, STS and from the test item. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 12 of 54 . The use of isolation gloves is highly recommended. . Perform all connections to the test item before powering-on the test set. . At the end of tests, power-off and then disconnect the test set. Further safety instructions are provided in the manual, for each test to be performed. Before operating, read them carefully! 1. 2 SAFETY SYMBOLS The following symbols, located on the test set, are used to alert the operator about dangerous points. The symbol is related to dangerous input or outputs, and it is located on the TD 5000 connector panel, by the side of the current input sockets, of the HV connector, and of the voltage Booster connector. . Also this symbol is related to HV generation, or to HV danger. On the TD 5000 connectors’ panel, the symbol is located by the side of the HV connector. . The symbol yellow socket. is located on the connector’s panel, close to the green- 1.3 RISKY SITUATIONS Never leave connection cables connected, even if the output is unused. The TD 5000 HV connector is suited for HV isolation; instead, connection cables have at their ends some accessible clip-on clamps. It is true that HV generation is disconnected when not used; however, an error could always occur. Connect the cable to the test device before enabling the test. Here again, a mistake is always possible. DO NOT CONNECT THE TEST DEVICE WHEN THE TEST SET IS ON! If, as the test is started, you understand that a clamp is not properly closed, or that it has dropped down, power off the test set before correcting! MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 13 of 54 Remember that, during these tests, the HV transformer is referred to ground: this is why touching conductors is so dangerous. In all tests, the generation is performed after having pressed the START button. After the start, the test set generates a low voltage, and measures the burden impedance: in case of high burden, including short-circuit, the operation is stopped, and the user is alerted. After this, the test set generates the full voltage, taking into account the voltage drop, so that the actual voltage copes with the nominal test value. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 14 of 54 2 TEST SET DESCRIPTION AND USE In this chapter we describe TD 5000, and inform about the HV generator characteristic. 2.1 THE FRONT AND SIDE PANELS The following pictures show the test set the side panels. All components are marked; their function is explained in the followings. In Appendix 1 are listed the standard components which are provided with the test set. For each component it is described its characteristic, and the mode of use. (1) TD 5000 ground connection (2) High voltage transformer connector, must be connected to the EXT. BOOSTER connector of the STS unit (3) STS communication port (4) “IN A” measurement input (5) “IN B” measurement input (6) Ground connection for the HV cable external shield (7) HV generator’s connector MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 15 of 54 2.2 LOCATING TD 5000 TD 5000, for the ease of transportation, is provided with front handles. It is conceived and tested to be operated when mounted on the foldable trolley, as shown in the following series of pictures. NOTE. If the trolley is not available, locate horizontally STS, and vertically TD 5000. Leave enough space around the test sets, so they are not heated by other devices. 2.2.1 Mounting the TD 5000 module There are two arms below: open them then put on TD 5000. After this, tie TD 5000 to the trolley with the stripe provided. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 16 of 54 2.2.2 Mounting the STS module Open the upper arms of the trolley: they host STS. Put on STS and tie it with the stripe. 2.2.3 Mounting the cable wheels The HV cable wheel is located in the rear, and blocked with the mechanical lock. The measurement cables are laid in the baskets. Now, for the ease of transportation, lift the handle, removing the mechanical lock: the trolley is ready to be moved around. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 17 of 54 2.2.4 Preparing for testing: modules connections When you are at the test place, first lower the handle; then, connect the communication cable. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 18 of 54 The Communication Port is a 15 pins connector. The housing is provided with two screws: after inserting the cable, firmly tighten them. Next, connect the BOOSTER cable. The BOOSTER connector is a safety one: after fitting it in, you have to turn it counter clockwise, until it stops: the mechanical lock prevents any unwanted disconnection. To disconnect it, you have to slide the lock down and rotate clockwise. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 19 of 54 Next, connect the ground cables. Connections are as follows: The STS module is connected to the substation ground, by means of the 6 m long cable provided, and to the TD 5000 module, with the short cable provided; The TD 5000 module is connected to the STS module and to the trolley, with the short cable provided. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 20 of 54 2.2.6 Preparing for testing: test cables connection The test set comes with the following clamps. The four ones to the left are used for measurements, with a smaller (above) or bigger (below) clamp opening, as a function of the thickness of the part you are clamping it on. The two to the right, with the 6 mm connector, are for the HV connection. Measurement cables are screwed into the corresponding connectors as shown. There can be one or two connections, according to the type of test to be performed. On the other side, the measurement cable is connected to one of the measurement clamps. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 The HV cable comes with a protection on the HV connector. The protection has to be unscrewed. Pag. 21 of 54 MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 22 of 54 Now, it is possible to screw the HV connector to TD 5000. The following picture shows the connector position when it is locked. For safety reasons, don’t forget connecting to ground the external shield! The connector has a mechanical protection: when fitting it, you will hear a soft click when the protection is armed. In this situation, the cable does not open even if you pull it: this is important, because otherwise the cable weight could cause the opening. To the other end of the HV cable, connect one of the two HV clamps: To remove the cable, push it some more, and then pull it open. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 23 of 54 2.2.7 Preparing for testing: connection to the power transformer A fundamental general rule is that the power transformer should be disconnected from the substation, and grounded before to start with test connections. In this example is shown how to prepare the setup for CH measurement of the following 400 MVA autotransformer. The connections show how to perform a GST measurement. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 24 of 54 First, the test set has been connected to ground with the tank. Next, all bushings must be short-circuited together and then connect the HV cable to one of the primary side bushing. The use of STS software to execute the measurement is described in section 3.1. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 25 of 54 2.3 POWER SUPPLY The power supply of the HV transformer is taken from the BOOSTER connections, and the power supply for the electronic boards is taken from the communication port. Attention: THESE CONNECTIONS MUST BE DONE WHEN STS IS POWERED OFF! Characteristics of the BOOSTER connection: Output not isolated from the mains supply. Output voltage: adjusted by STS from 0 to 220 V AC, as a function of the test being developed. Output power; STS supply 230 V: 1500 VA steady, 3600 VA during 2 minutes. Output power; STS supply 110 V: 1360 VA steady, 2500 VA during 2 minutes. The Ground connection is critical for some of the tests to be performed, and is the main protection against errors. For this reason, TD 5000 is provided with a dedicated ground connection cable, which shall be connected to ground as close as possible to the test power transformer. STS will be connected to the same ground by means of the short yellow/green cable provided. THE TEST SET CANNOT BE OPERATED IF NOT CONNECTED TO GROUND: THIS PREVENTS POSSIBLE DANGERS TO THE OPERATOR AND/OR FAULTS OF THE TEST SET. 2.4 POWER-ON When you are to operate the TD 5000 unit, before power-on, connect STS to TD 5000, as explained here before. As you power-on STS, also TD 5000 is poweredon. As soon as STS is connected to the mains, the power supply goes to a standby mode, a diagnostic runs, and, in a couple of seconds, the test set can be powered-on. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 26 of 54 Pressing the ON/OFF button powers-on the test set; pressing it again, the test set is turned off. After power-on, the test set will perform first a self-diagnostic sequence, and then the display will confirm that TD 5000 is connected. The TD 5000 connection status with STS is shown with the following icons on the main window: TD 5000 is communicating properly with STS TD 5000 is not communicating or it’s not connected TD 5000 is communicating but not properly 2.5 GENERATOR CHARACTERISTICS The TD 5000 module, driven by STS, is a High Voltage electronic generator, with the following characteristics, in the range of 50 to 60 Hz. Generator characteristics are the followings. MAX. OUTPUT VOLTAGE 12000V 12000V OUTPUT CURRENT 300 mA 125 mA MAX. OUTPUT DURATION >120 s >1h OFF/ON RATIO 6.3 1.1 The ON/OFF ratio is the ratio of the time during which the generation is blocked, after having generated high power. When the current generation is less than the maximum output duration, the OFF time is the ON time by the OFF/ON ratio. For instance, if the test at 300 mA lasted 20 s, the OFF time lasts: TOFF = 20 * 6.3 = 126 s. On this output there is no output or power limitation as a function of the supply voltage. The HV output decreases in amplitude below 45 Hz. The decrease is linear with the frequency; at 15 Hz the amplitude is 40% of the one at 50 Hz. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 27 of 54 The HV output decreases also for frequencies above 200 Hz, but only at 300 mA. The following diagram shows the test set behaviour. 120 100 [%] 80 60 < 125 mA 40 < 300 mA 20 15 40 65 90 115 140 165 190 215 240 265 290 315 340 365 390 0 Frequency [Hz] The test set measures the current generated by the HV output, and the phase shift of the current with respect to the voltage. These measurements don’t change with the HV output range. There are two current measurement ranges. Current ranges, resolution and accuracy are shown in the next table. INTERNAL MEASURE RESOL. 1V 1 mA TYPICAL ACCURACY ± % (rdg) ± % (rg) ± 0.2% ± 0.5 V ± 0.2% ± 1 mA GUARANTEED ACCURACY ± % (rdg) ± % (rg) ± 0.3% ± 1 V ± 0.5% ± 2 mA 12000 V AC 5 A AC (@ inputs A or B) 10 mA AC (@ inputs A or B) 0.1 µA ± 0.2% ± 0.1 µA ± 0.3% ± 0.1 µA NOTE. When we speak of accuracy, the error has always two components: the first one proportional to the reading (rdg), the second one proportional to the range (rg). MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 28 of 54 Frequency range: 15 to 500 Hz. Frequency resolution: 10 mHz; accuracy 10 ppM. Connections: by a double shielded HV connector, a Ground sockets, and two measurement sockets (A and B). From the measurements of V and I, the following measurements are derived. Capacitance: Measurement range 1, from 1 pF to 100nF. Resolution: 6 digits. Accuracy, typical: ± 0.03% of the value ± 0.1 pF; guaranteed: ± 0.05% of the value ± 0.1pF. Measurement range 2, from 10 nF to 3 µF. Resolution: 6 digits; accuracy, typical: ± 0.1% of the value ± 10 pF; guaranteed: ± 0.2% of the value ± 10 pF. Tan(δ)(or dissipation factor DF): Measurement range 1: from 0 to 10% (capacitive). Resolution: 5 digits; accuracy, typical: 0.05% of the value ± 0.005 %; guaranteed: 0.1% of the value ± 0.005 %; Measurement range 2: from 0 to 100%. Resolution: 5 digits; accuracy, typical: 0.3% of the value ± 0.01 %; guaranteed: 0.5% of the value ± 0.02 %. Measurement range 3: over 100%. Resolution: 5 digits; accuracy, typical: 0.5% of the value ± 0.03 %; guaranteed: 0.8% of the value ± 0.05 %. Power factor PF (or cos(φ)): Measurement range 1: from 0 to 10% (capacitive). Resolution: 5 digits; accuracy, typical: 0.05% of the value ± 0.005 %; guaranteed: 0.1% of the value ± 0.005 %; Measurement range 2: from 0 to 100%. Resolution: 5 digits; accuracy, typical: 0.3% of the value ± 0.02 %; guaranteed: 0.5% of the value ± 0.02 %. Power: Measurement ranges: 10 kW, 100 kW, 1 MW. Resolution: 0.1 mW; accuracy, typical: 0.5% of the value ± 1 mW; guaranteed: 1% of the value ± 2 mW. Inductance: MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 29 of 54 Measurement range 1: from 1 H to 10 kH. Resolution: 0.1 mH; accuracy, typical: 0.5% of the value ± 0.5 mH; guaranteed: 1% of the value ± 1 mH. Measurement range 2: from 100 H to 1 MH. Resolution: 1 H; accuracy, typical:0.5% of the value ± 0.5 H; guaranteed: 1% of the value ± 1 H; MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 30 of 54 3 TAN(δ) AND CAPACITANCE MEASUREMENT BASICS The electrical model of a generic test object can be modeled as a capacitor with a parallel resistor as shown in the following picture: IC ITOT C IC R IR ITOT d IR Applying an AC voltage to the test object, the total current flowing out from the voltage source is divided into the capacitive and resistive part of the impedance. The tan(δ) quantity, hereafter called TD, is defined as: The parallel resistor represents how much the capacitor is not ideal, in other words it is an indication of the dielectric losses: as small is the resistance, as bad is the dielectric. Accordingly to this concept, as good is the dielectric, as small is TD. The TD measurement is used, for example, to understand how good is the insulation of a bushing or the insulation between the HV and LV winding of a power transformer. Since the resistive part of an insulator is very high, a very high voltage source with very high power must be used in order to drive such a huge resistive load and inject a small measurable current. NOTE: TD can be called Dissipation Factor (DF), these two terms are considered synonymous. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 31 of 54 From the two currents IC and IR is possible to derive another quantity called Power Factor, hereafter defined as PF: These two quantities have practically the same value when d is very close to zero; if your test object has a very good insulation who leads to a very small TD value, even PF has the same small value. But if the test object has a problem and TD starts to increase, beyond few degrees of d angle, TD and PF starts to be slightly different. TD value should not change if the applied voltage changes; actually what’s happen is that beyond a certain voltage level, the applied electrical field starts to polarize the insulator’s atoms so the conductivity changes as well. This atoms polarization effect leads to an increasing of TD values even if the insulator is not damaged. Different materials give different response to the same voltage. Even changing the applied frequency the TD values change, and the main reason is that the capacitor impedance is a function of frequency. TD [%] TD [%] 1.5 15.0 1.0 10.0 0.5 5.0 1k 5k 10k [V] 15 200 400 [Hz] All physical phenomenon who interact and play a role in the TD value change are various and their physical description and comprehension is complicated, so the concept “as good is the dielectric, as small is TD” automatically lose of sense if it is not related to the test conditions (voltage, frequency, temperature, etc…) and to what kind of insulation (oil, paper, resin, porcelain, etc…) we are talking about. What must be very clear to the operator is that a single value of TD or PF alone cannot be used to state whether the insulation is good or damaged, what has sense to do is to collect many values and do a comparison in order to see if there is a worsening of the TD values’ trend over the time. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 32 of 54 3.1 CAPACITANCE AND TD MEASUREMENT OF A GENERIC TEST OBJECT In the STS main window, select the “Tangent Delta & Capacitance” icon (see picture above) and press the knob. The following windows will be opened: (1) Test type. Select here where the high voltage output will be applied: in case of a defined test object (CT, VT, PT or CB), the available options will be displayed accordingly. (2) Capacitance under test. In case of e defined test object (CT, VT, PT or CB), the available options will be displayed accordingly. (3) Generation mode. Select here if a single shot, a voltage sweep or a frequency sweep must be performed. (4) Test Mode. Select here the test mode accordingly with the connection between TD 5000 unit and the test object. In case of a defined test object MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 33 of 54 (CT, VT, PT or CB), the most proper test mode will be automatically selected accordingly with the selections (1) and (2). See section 4 for further details. (5) Voltage/Frequency test table. Accordingly with selection (3), this table allows to set different test voltages or frequencies. (6) Nominal Values. Insert here the reference values of capacitance and TD. In case of e defined test object (CT, VT, PT or CB), these values will be taken from the related headers (see MIE12175 – STS FAMILY Application Guide). The check-box allows including these parameters in Tests table. (7) Temperature compensation. Capacitance and TD values changes with the temperature: selecting the check-box, the “k” coefficient will be used to compensate the measurements. (8) Tests table. To add a test in the test table, highlight this table and press the button “Add Test”; each row of this table contains the parameters selected in (1) (2) (3) (4) (5) (6) (7). (9) Results table. Here you find all measurements and calculations of the selected row in the Tests table. (10) Show graph. In case of a voltage or a frequency sweep, capacitance and TD values are plotted in a graph. 3.1.1 Tests execution Connect the TD 5000 unit to STS while STS is switched off, then switch on STS; the TD 5000 unit is powered by STS. Prepare all tests you want to perform and add rows in the Tests table (see section 3.1); in the following picture is shown an example of a voltage sweep: MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 34 of 54 Connect TD 5000 to the test object taking care of the selected test mode (e.g. UST-A, GSTg-a, and so on). Once all connections are done and all safety precautions has been respected, press the START/STOP button . Each pair of voltage and frequency injection consists in a sequence of operations: 1. 2. 3. Load evaluation Calibration Measurement This sequence is performed in order to guarantee the maximum possible measurement accuracy. 3.1.2 View Results Once all tests are performed, the Results table contains all measurements. The Results table changes every time a row of the Tests table is highlighted and selected by pressing the knob. When a row in Tests table is selected, the focus is automatically moved on the Results table and the window is displayed as in the following picture: MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Press the button “Scroll” table. Pag. 35 of 54 and rotate the knob to scroll left-right the Results With the button “Show graph” , in the Results table will be shown the corresponding graph of the measurements. The following pictures are examples of capacitance and TD measured with a frequency sweep: MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 36 of 54 Rotate the knob to highlight the graph and then press it: with this operation it is possible to read the values of each point: 3.1.3 Save results Press the save button located to the left of the screen. For more details, please refer to MIE12175 STS FAMILY – Application Guide. 3.1.4 Open results Press the open button located to the left of the screen. For more details, please refer to MIE12175 STS FAMILY – Application Guide. 3.1.5 Software settings MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 37 of 54 Selecting the two icons above and the panel “Tandelta”, the following window will be shown: Power factor naming. Select if you want to consider TD or PF values. Power factor value. Select if TD or PF values must be shown as an absolute value or a percentage value (% value = abs value * 100). Noise Reduction. Select how many measures will be taken to perform an averaging and calculate the result. Ground Shield Check. Enable the sensing of the HV shield cable’s connection. Warnings. Enable external warnings devices during the execution of the test: buzzer, siren or strobe light. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 38 of 54 4 TEST MODES There are two main test modes: GST - Ground Specimen Test UST - Ungrounded Specimen Test The target is to measure the current that is flowing in a capacitor in order to perform capacitance, tangent delta or power factor measurement. When the test object is, for example, a two windings power transformer, the equivalent circuit is composed by three interconnected capacitors, thus the total current flowing out from TD 5000 will be divided into each branch of the circuit. The only way to perform the measurement on a single capacitor is to filter out (or to guard) the current that is flowing into the other branches. Accordingly with the selected test mode, TD 5000 will automatically set-up the internal architecture in order to guard the unwanted currents. The following picture shows this principle on a two windings power transformer. ITOT = I1 + I2 CHL TD 5000 HV output I2 I2 CH I1 CL X GND IN A IN B SWITCH MATRIX Current sensing I1 I1 I2 CH measurement, test mode GSTg-A If the capacitance CH must be measured, the only current that must be taken into account is I1. Since CH cannot be disconnected from CHL and CL, the current I2 must circulate out of the current sensing unit; through the switch matrix a different path is created and I2 is deviated to point “X”. Note that any current can flow into CL because IN A creates a very low impedance node, so I2 is not divided. The test mode here described is called GSTg-A. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 39 of 54 With the same connection is possible to measure CHL and the sum CH+CHL, just selecting different test modes, UST-A and GST respectively (see pictures below). In case of UST-A the current that must be guarded is the one that is flowing in the ground connection, while in case of GST all currents must be considered. ITOT = I1 + I2 CHL TD 5000 HV output I2 I1 CH I1 CL X GND IN A IN B SWITCH MATRIX Current sensing I1 I2 I2 CHL measurement, test mode UST-A ITOT = I1 + I2 CHL TD 5000 HV output I2 CH I1 CL X GND IN A IN B SWITCH MATRIX Current sensing I1+I2 I1 I2 CH+CHL measurement, test mode GST To measure CL a different connection is needed: MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 40 of 54 ITOT = I1 + I2 CHL TD 5000 HV output I2 I2 I1 CH CL X GND IN A IN B I2 Current sensing SWITCH MATRIX I1 I1 CL measurement, test mode GSTg-A Same as CH, the capacitance CL can be measured with the GSTg-A test mode. NOTE: the same tests can be performed using IN B instead of IN A as well, just select test mode GSTg-B for CH and CL and UST-B for CHL. The most complicated case is the three windings power transformer. The third winding introduces three more equivalent capacitances: CT, CLT and CHT. In the following picture is shown that both IN A and IN B must be used together in order to measure CH: in this case only GSTg-A+B test mode can be used. ITOT = I1+I2+I3 CHT I3 TD 5000 HV output CLT CHL I2+I3 X GND IN A IN B SWITCH MATRIX Current sensing I2 I1 CL CH I1 I2 I3 CH measurement, test mode GSTg-A+B CT MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 41 of 54 5 POWER TRANSFORMERS Select this icon on the STS main window, input all necessary header parameters and select the “Tangent d” button in the test list. The use of STS software to execute the measurement is described in section 3.1. 5.1 TWO WINDINGS TRANSFORMER Before to start with any measurements the transformer must be taken out of service, disconnected from the power system and the tank must be very well grounded. All HV bushings must be short-circuited, as well as the LV bushings. Once these operations are done, three equivalent capacitances must be measured (see picture below): CH: capacitance between HV winding and the ground CHL : capacitance between HV and LV winding CL: capacitance between LV winding and the ground CHL CH CL MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 42 of 54 5.1.1 Test connections 5.1.1.1 Capacitance and TD measurement of CH, CHL or CH+CHL HV cable's shield To STS EXT. DEVICE port CHL CH CL To STS External Booster socket To STS ground 5.1.1.2 Capacitance and TD measurement of CL, CHL or CL+CHL To STS EXT. DEVICE port HV cable's shield CHL CH To STS External Booster socket To STS ground CL MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 43 of 54 5.2 THREE WINDINGS TRANSFORMER Before to start with any measurements the transformer must be taken out of service, disconnected from the power system and the tank must be very well grounded. All HV bushings must be short-circuited, as well as the LV and tertiary bushings. The presence the tertiary winding complicates the equivalent circuits. In this case there are six capacitances to be measured: CH: capacitance between HV winding and the ground CHL : capacitance between HV and LV winding CL: capacitance between LV winding and the ground CLT: capacitance between LV and tertiary winding CT : capacitance between tertiary winding and the ground CHT: capacitance between HV and tertiary winding 5.2.1 Test connections 4.2.1.1 Capacitance and TD measurement of CH, CHL, CH+CHL, CHT, CH+CHT, CHL+CHT or CH+CHL+CHT HV cable's shield To STS EXT. DEVICE port CHT CHL To STS External Booster socket To STS ground CH CLT CL CT MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 44 of 54 5.2.1.2 Capacitance and TD measurement of CL, CHL, CL+CHL, CLT, CL+CLT, CHL+CLT or CL+CHL+CHT HV cable's shield To STS EXT. DEVICE port CHT CHL To STS External Booster socket CLT CH CL CT To STS ground 5.2.1.3 Capacitance and TD measurement of CT, CHT, CT+CHT, CLT, CT+CLT, CHT+CLT or CT+CHL+CHT To STS EXT. DEVICE port HV cable's shield CHT CHL To STS External Booster socket To STS ground CH CLT CL CT MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 45 of 54 5.3 AUTOTRANSFORMER Before to start with any measurements the transformer must be taken out of service, disconnected from the power system and the tank must be very well grounded. Since the autotransformer has a single winding, all HV and LV bushings must be short-circuited together. Once these operations are done, there is only one equivalent capacitance: CH: capacitance between HV/LV winding and the ground 5.3.1 Test connections 5.3.1.1 Capacitance and TD measurement of CH HV cable's shield To STS External Booster socket To STS ground CH MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 46 of 54 5.4 AUTOTRANSFORMER WITH TERTIARY Before starting with any measurements the transformer must be taken out of service, disconnected from the power system and the tank must be very well grounded. All HV and LV bushings must be short-circuited together, as well as the tertiary bushings. This power transformer has two windings, so the equivalent circuit is the same of a two windings transformer; the capacitances to be measured are: CH: capacitance between HV/LV winding and the ground CHT : capacitance between HV/LV and tertiary winding CT: capacitance between tertiary winding and the ground 5.4.1 Test connections 5.4.1.1 Capacitance and TD measurement of CH, CHT or CH+CHT HV cable's shield To STS EXT. DEVICE port CHT CH To STS External Booster socket To STS ground CT MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 47 of 54 5.4.1.2 Capacitance and TD measurement of CT, CHT or CH+CHT HV cable's shield To STS EXT. DEVICE port CHT CH To STS External Booster socket To STS ground CT MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 48 of 54 5.5 BUSHING A bushing has two equivalent capacitances to be measured: C1: capacitance between the internal conductor and the Td tap C2: capacitance between the Td tap and the ground 5.5.1 Test connections, capacitance and TD measurement of C1 HV cable's shield CONDUCTOR C1 To STS EXT. DEVICE port To STS External Booster socket To STS ground C2 Td TAP MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 49 of 54 5.5.2 Test connections, capacitance and TD measurement of C2 To STS EXT. DEVICE port CONDUCTOR HV cable's shield C1 To STS External Booster socket C2 Td TAP To STS ground 5.5.3 Hot collar test For a correct use of the hot collar, the inner black side of the collar must be placed directly in contact on the surface of the bushing under test. The green colored other side of the collar is not so conductive; therefore the outer green colored side is not a so suitable for an electrical contact. In the next picture there is an example of a correct application of the collar. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 50 of 54 As you can see, the inner black side of the belt is strictly wrapped around the device under test. In the picture the Device Under Test (D.U.T) is the lower part of the bushing of a medium voltage breaker. The HV is applied via the HV clamp; be sure that a good electrical contact is present between the clamp and the black surface of the collar. In the ISA's supplied hot collars, a metallic buckle is present to ensure a good tensile force as well a good electrical contact. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 51 of 54 Here above an example of a hot collar wrapped and fastened with the buckle. In the following images, a sequence of the operations required for the hot collar installation. 1) Wrap the collar around the bushing; then insert the free side of the collar into the buckle as shown. Take some care to place the small part of the belt which is “solid” with the buckle; place it just under the free part of the belt. This will ensure a continuous contact all around the surface of the D.U.T and the conductive side of the collar itself. 2) Once the operations above are executed, fasten and lock the belt into the buckle. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 52 of 54 3) If there’s enough free space around the wrapped collar; you can engage the clamp directly on the buckle. Otherwise it is possible to hang the clamp to the belt directly. In this case, be sure that the black side of the collar belt is in electrical contact with the metallic part of the clamp. In both cases avoid any contact of the not wrapped part of the collar with the objects around. Don't leave it “floating” around. Wrap the remaining part on itself. MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 53 of 54 Of course, it is possible connect the HV using the 4 mm plug connected on the collar as in the following image If erroneously the collar is wrapped “upside down” (see below), that means with the colored side attached to the DUT surface, you will be able anyway to perform the test, but the measurements will be affected by this fact (see table below, where are compared the results on the same D.U.T). Results using the collar installed in the right way: Vout [V] Iout [A] Freq. [Hz] Cp [F] Tδ QF Loss [W] [W]@10k QTest [VA] STest V [VA] Rp [Ω] 1 4992.4 24.4µ 35.0 22.2108p 8.5218m 117.3465 1.0370m 4.1606m -0.1217 0.1217 24.0349G 2 4984.4 34.7µ 50.0 22.1851p 4.9392m 202.6650 0.8533m 3.4343m -0.1731 0.1731 29.1485G 3 5007.9 94.3µ 135.0 22.1732p 4.2812m 233.5770 2.0208m 8.0578m -0.4720 0.4720 12.4104G 4 5008.4 0.2m 220.0 22.1804p 4.8315m 206.9737 3.7111m 14.7943m -0.7681 0.7681 6759.4389M 5 4983.4 0.2m 305.0 22.1535p 5.9287m 168.6721 6.2559m 25.1906m -1.0552 1.0552 3969.7322M MIE13175 TD 5000 – Application Guide – Rev 1.1.1 Pag. 54 of 54 Results using the collar “reversed” : Vout [V] Iout [A] Freq. [Hz] Cp [F] Tδ QF Loss [W] [W]@10kV QTest [VA] STest [VA] Rp [Ω] 1 4994.4 23.3µ 35.0 21.1884p 2.9884m 334.6243 0.3472m 1.3919m -0.1162 0.1162 71.8720G 2 4987.2 33.2µ 50.0 21.1802p 1.9175m 521.6025 0.3170m 1.2743m -0.1654 0.1654 78.6833G 3 4990.0 89.7µ 135.0 21.1877p 1.6289m 613.9007 0.7295m 2.9296m -0.4478 0.4478 34.1361G 4 5008.5 0.1m 220.0 21.2090p 2.2410m 446.2352 1.6459m 6.5614m -0.7345 0.7345 15.2424G 5 5001.8 0.2m 305.0 21.1932p 2.5988m 384.7918 2.6428m 10.5636m -1.0169 1.0169 9466.6862M 6 4987.1 0.3m 400.0 21.1982p 2.8558m 350.1664 3.7829m 15.2102m -1.3247 1.3247 6574.7039M
© Copyright 2024 ExpyDoc