EPM 4 - data sheet

Electrosurgery
Analyzers
EPM 4
Product-Overview:
EPM 4 analyzes Electro-Surgical Units quickly and accurately. A wide load-resistance
range provides user-selectable loads, including very low loads for testing many of
today’s ESU’s.
The first measuring cabinet in the world of the new generation electrical operating
tools for continuous recording of pulsed currents.
The automatic calibration of the RMS current transducer every 30 sec guarantee an
exact test unit.
Product-Highlights:
The first measureing cabinet in the world of the new generation electrical operating
tools for continuous recordings of pulsed currents.

Measurement of capacity curves

Current measurements:0,01 Arms (0,014 A) up to 6 Arms (8,6 A), protected
from overload

Two large load resistance ranges with fine graduations:
10 , 25  …6375  in 25  - intervalls
2,5  ……….43,75  in 6,25 -  intervalls

Additional fixed resistor:
200  (10 k optional)

Vertex factor (crest factor)

110 mm * 60 mm LCD

Automatic calibration of the RMS current transducer every 30 seconds

Wide range power adapter for global use (90 – 260 V / 50 – 60 Hz)

Windows-Software for testing with a protocoll in accordance with instruments
from the manufacturer

Scope of delivery EPM 4: Plant calibration using the measureing protocol

Adapter cable sets for diverse operating tools are optional
EPM 4 - Technical Data
Current Measurement
0,01 Arms … 6 Arms (0,014 Ap
… 8,6 Ap)
Overload protected
Current monitor
(1 Ap = 1 Vp)
Attention:
We claim the right to alter
specifications without notice.
2 Loadresistor Ranges

10 , 25 … 6375,00 

2,5 … 43,75 
Additional fixed-resistor
200  (10 k optional)
Crest-Factor
110 x 60 mm LC-Display
Automatic calibration of the
RMS-current transducer every
30 seconds
Wide range mains connection
unit for global use
(90 – 260 V / 50 – 60 Hz)
Manufacturer calibration with
certificate
Single measurement
(SINGLE)
Single triggering and storage of
power, current, crest-factor and
resistance
Continous measurement
(CONT)
Continuous indication of power,
current, crest-factor and
resistance for adjustments
Automatic recording of power
distribution
(POWER DIST)
Automatic measurement of
loadpoint sequence
(LIST)
Recording of (max. 10 s)
pulsed currents
Tabular display of pulse
duration and power
Graphs scaleable
02/2010
Presentations and more information:
Manufacturer:
Nordring 67, 63843 Niedernberg
Telefon (06028) 999622-0 Fax:999622-11
E-Mail: [email protected]