Manual

How to use Symphonix with The P7889 Acquisition board for
fast acquisitions
The P7889 hardware is controlled by a software server, and Symphonix has been modified to
dialog with this server to control the acquisition board and retrieve data. Then to use Symphonix
with this acquisition board, before running Symphonix, P7889 server must be started and properly
configured, otherwise Symphonix will prompt for an error at start-up. To start P7889 server,
double click on Windows desktop P7889 icon. Server starts and MCDWIN, fastcom software
opens.
To load the proper hardware configuration file, In MCDWIN click on "range preset" icon
The P7889 settings windows is displayed
In setup Name edit type inel . (Pay attention this is case sensitive).
Next click on load Sett button follow by apply button. This will load and apply the proper settings
to use the P7889 with an Inel CPS detector.
From now, the acquisition board server is configured and MCDWIN may be closed
Note : P7889 server must be always kept on running
Using Symphonix
Start Inelservers (desktop icon). Two others animated icons should be in the icons taskbar on the
bottom right corner (on the Window desktop). Depending on the acquisition mode used with the
P7889 board, 2 modules have been implemented in Symphonix. One is called acqpci64 standard
for classical use of the acquisition board and another one, called AcqPci64 sequence to proceed
to very fast cyclic acquisition runs.
Acquisition mode selection
To switch form one mode to another one, you have to open Inelserver. For this, double click on
the animated Inelservers icon. Then the about servers window will popup.
To do classical acquisitions, thus use acqpci64 standart mode, in COM-Acquisition mode selector
Combobox must be on AcqPci64.AcqPci64Object
To run acquisition sequences, acqpci64 sequence the same Combobox must be on
AcqPci64Sequence.AcqPci64SequenceObject
To check what is the current selected mode, set the mouse over (above) detector running icon,
wait few second until the hint message appears. This message indicates the current acquisition
mode used.
Once the desired working mode has been selected, press the Close button to close this window.
mode "acqpci64 standart"
With this mode , acqcps.exe may be used. (launching the 'Acqcps” icon on desktop)
mode "acqpci64 sequence"
With this mode, only the "sequences" window must be used.
To open it, click on "acqpci64 sequence" running detector icon.
Select the desirable spectra number (number of different acquisitions in a cycle), cycles number
(number of cycles to do), acquisition preset (acquisition time for one acquisition during one cycle)
and then click on Start.
A Window will popup to ask you to click on apply in the p7889 windows before click on the Ok
button of this Window.
The acquisition sequence is started.
When finished, an end of Acqs message appears. To see the different diagrams, you can select
the spectrum number to be displayed in the current showing spectra box.
To save all diagrams in FDT format file click on save fdt button. The fdt filename appears under
and may be changed using light blue button on the right.
Known issues :
AcqPci64 will raise the exception “catastrophic failure” if the Inelserver is started but not P7889
server.
Launch P7889 and then restart AcqPci64.