RNA Medical CVC 9005

RNA CVC 9005
Calibration Verification Controls
How to Locate a Calibration Control Lot
1.
Select Bookmark tab on left:
2.
Locate RNA Kit #:
• Select appropriate kit
• If using the H4 Cartridge, please note sheet specific for H4
3.
To Find a Specific Level and Lot:
• Open file by selecting the “+” icon
• The expanded list of control levels will display
IRMA TRUpoint®
Calibration Verification Control
For use with the RNA Medical, Inc.
C V C 9005 C a l i b r a t i o n V e r i f i c a t i o n C o n t r o l s
RNA Medical CVC 9005
Calibration Verification
Control
Kit 3900509
Level 1 Lot 37509
Level 2 Lot 37609
Level 3 Lot 37709
Level 4 Lot 37809
Level 5 Lot 37910
EXP. 06/15
For use with CC, BG, and H3 Cartridges
Expected Values Chart *
LEVEL
1
2
3
4
5
LOT
NUMBER
37509
37609
37709
37809
37910
HCT (%)
15.4 – 21.4
20.6 – 26.6
30.9 – 36.9
43.3 – 49.3
56.7 – 62.7
The values on the Expected Values Chart are based on the results of multiple
sample analysis. Samples were drawn with a 1mL syringe, 18 gauge plastic
needle tip, after equilibration at 22°C (72°F) for a minimum of 8 hours.
Use the expected values for each parameter as a guide in evaluating
performance. Since performance is subject to sample temperature and
environmental barometric pressure, ITC recommends that each institution
establish their own expected values and acceptable limits. The mean values
established at your institution should fall within the expected ranges.
*NOTE: These values are for use with the IRMA TRUpoint only.
CVC9005_CC_3900509_0615
IRMA TRUpoint®
Calibration Verification Control
For use with the RNA Medical, Inc.
C V C 9005 C a l i b r a t i o n V e r i f i c a t i o n C o n t r o l s
RNA Medical CVC 9005
Calibration Verification
Control
Kit 3900509
Level 1 Lot 37509
Level 2 Lot 37609
Level 3 Lot 37709
Level 4 Lot 37809
Level 5 Lot 37910
EXP. 06/15
For use with H4 Cartridges
Expected Values Chart *
LEVEL
1
2
3
4
5
LOT
NUMBER
37509
37609
37709
37809
37910
HCT (%)
16.9 – 22.9
21.3 – 27.3
31.7 – 37.7
43.6 – 49.6
66.4 – 72.4
The values on the Expected Values Chart are based on the results of multiple
sample analysis. Samples were drawn with a 1mL syringe, 18 gauge plastic
needle tip, after equilibration at 22°C (72°F) for a minimum of 8 hours.
Use the expected values for each parameter as a guide in evaluating
performance. Since performance is subject to sample temperature and
environmental barometric pressure, ITC recommends that each institution
establish their own expected values and acceptable limits. The mean values
established at your institution should fall within the expected ranges.
*NOTE: These values are for use with the IRMA TRUpoint only.
CVC9005_H4_3900509_0615