Elektronische Temperaturmesstechnik | LABOM GmbH

Electronic digital thermometer
Type series GE4 . . .
Features
■ Electronic digital thermometer
■ Case and wetted parts of stainless steel
■ Degree of protection IP 67
■ Quick repsonse time with lasting accuracy
■ Easily read display with either analouge or digital
indication
■ Sturdy mechanical design and construction make it
suitable for numerous applications
■ Completely maintenance free for the life span of the
product, by means of a solar cell or a battery (life
span ten years)
■ No mechanical wear (or ageing)
Application
Currently, for industrial applications, local indicating thermometers are available which work without auxiliary power but
which only have an analogue dial available. They are filled with
mercury or gas or are constructed as bimetallic thermometers.
On account of their extremely low energy consumption in
conjunction with a nearly loss-free operating microprocessor,
the electronic digital thermometers from LABOM allow an
entirely novel measuring device to be realised.
The electronic digital thermometer is equipped with a lithium battery as its energy source, and this has an operating
lifetime of 10 years. Furthermore, remote thermometers are
available. They are easily installed because connection to the
sensor and the display unit is made by means of individually
adapted connection cables.
Techn. Data
Case
stainless steel mat. no. 1.4301, DN100
Window
non splintering laminated glass
Accuracy
per EN 13190, class 1
Process connection
radial bottom or centre back connection or
with capillary.
Different connections available
Case seal
Perbunan
Storage temperature
max. -20 ...+60 °C, other values upon request
Case design
degree of protection IP 67 per EN 60529,
Measuring resistor
Pt 1000
Nominal ranges
per EN 13190 up -60...+500 °C
Dual display
· analogue:
· digital:
bar graph 61 divisions
4 digit, 7 segment display
11 mm high
Power supply
Lithium battery
Service life
10 years
Temperature detecting element
stainless steel mat.-no. 1.4571, Ø 8 mm,
standard lengths available
Data Sheet: T4-042 Rev. 0D7
Data Sheet: T4-042 Rev. 0D7
The measuring process
The Electronic digital thermometers rely on a periodic measurement
process. The cycle time is 3 seconds, so that the viewer, even in
passing, can take a current reading. An activity display (a sun symbol
at the start of the display) indicates the measurement cycle.
The display
The digital thermometer provide a dual display of LCD elements.
The original concept for the format of the display was to appear
like a traditional dial instrument. A circular arrangement of LCD
elements achieves this. This quasi-analogue display allows a rapid
interpretation even at extended distances. The exact temperature
can be read from the 4 digit segment display. Depending on the
range programmed, the display can have a resolution of either
0.1 K or 0.5 K.
The constructive design
The electronic thermometer consists of a robust assembly in a
stainless steel case. To accommodate industrial applications the
unit complies the vibration requirements of German Lloyd. The
type of proctection is to IP 67 and the electromagnetic interference
complies with the requirements of the EC standards.
Sensor connections
The temperature sensor is a Platinum resistance type Pt 1000. Its
signal conditioning is performed by the micro processor.
The small size of the sensor chip gives pinpoint positioning in the
Dimensions
medium by faster response characteristics. The indicator can be
positioned close to the sensor or at a distance of up to several
meters. Leads can be made to customers requirements. Unsightly
loops are then avoided on site.
Performance
The micro processor conditions the signal resistance of the
Pt 1000 with a resolution of 12 bits. The numerical error of
measurement is thus less than 1 Kelvin. This fulfills the accuracy
requirements for measuring equipment to class 1.
Self test
After being switched on the device performs an initialization routine. In the process it carries out a self test as well as checking the
operation of the sensor and the connections.
Sensor monitoring
The micro processor continually checks the plausibility of the sensor
signal. The result is displayed on the display.
Service considerations
A significant advantage of our instrument lies in the long term
stability of the use of electronics to provide the measurement and
by the lack of mechanical components.
Integrated diagnostic routines continually check the validity of
the sensor output so that a sensor failure can be flagged on the
display.
Data Sheet: T4-042 Rev. 0D7
Dimensions (cont.)
G
G 1/2 B
1/2" NPT
G 3/4 B
G1B
Ød2
26
32
39
Order Details
h1
12
20
12
16
sw
27
27
32
41
G
G 1/2
M24x1.5
t1
19
17
- please give additional specifications for models not listed -
Electronic digital thermometer
design
accuracy
· connection bottom
· connection back
· connection bottom
with capillary
· connection back
· standard class 1 (nominal range)
with rigid temperature
detecting element
-60...+40° C
-40...+80°C
0...+120°C
nominal range
0...+160°C
0...+250°C
0...+300°C ²
connection
· without screwing
temperature detecting element Ø · 8 mm
· 100 mm
· 160 mm
immersion length
· 250 mm
· 400 mm
· mm as in writing
·1m
·3m
length of capillary¹
·5m
· as in wrting
· for surface mounting (back flange)
mounting
· for flush mounting (front flange)
order code (example):
GE4100
GE4110
GE4300
GE4310
A2 . . .
155
223
540
544
560
565
D1001
F8 . . .
100
160
250
400
999
K31
K33
K35
K39
G2
G3
GE4100 A2155 D1307 F8100
¹ with design capillary, only.
² we recommend capillary design for ranges > 200 °C.
LABOM Mess- und Regeltechnik GmbH · P.O. Box 12 62 · 27795 Hude · Germany · Im Gewerbepark 13 · 27798 Hude · Germany
Tel. +49 4408 804-0 · Fax +49 4408 804-100 · e-mail: [email protected] · www.labom.com