Tender Ref. No: 23/E(P) /CGMSC/EQP/2014, Dt 22

CHHATTISGARH MEDICAL SERVICES CORPORATION LTD.
(Chhattisgarh Govt. Undertaking)
Amendment No-1, Date17/10/2014
Tender Ref. No: 23/E(P) /CGMSC/EQP/2014, Dt 22 /09/2014
Following amendments are being incorporated in the referred tender enquiry document.
S. No.
Particulars (as per existing tender)
1.
Last Date For Online Submission Of Tender:
22/10/2014
2.
ELIGIBILITY CRITERIA (Page No. 7)
2.1 Manufacturers with ISO 13485 certification are
eligible to participate in the tender provided:
(i) They have valid manufacturing license for item
2.1 quoted/proof as manufacturer of product quoted.
2.2
2.3
3.
3.1
3.2
3.3
3.4
3.5
3.6
4.
4.1
a.
b.
Amended
Last Date For Online Submission Of Tender:
31/10/2014
2.1 Manufacturers with ISO certification are
eligible to participate in the tender provided:
(i) They have valid manufacturing license for
item quoted/proof as manufacturer of product
quoted or any other document like USFDA /
ISO indicating type of item & mentioning
manufacturing site.
2.2 (ii) Average annual turnover of the bidder of
3 crore or more in the three (3) financial years,
as per Annexure V.
2.2 (iii) Should furnish copy of manufacturing
license& ISO certificate of the original
manufactures.
2.2 (ii) Average annual turnover of the bidder of 5
crore or more in the three (3) financial years, as
per Annexure V.
2.2 (iii) Should furnish copy of manufacturing
license& ISO 13485 certificate of the original
manufactures.
Annexure-II (Schedule of Requirement)
S. No.11 EQ2011: Electro cautory with
S. No.11 EQ2011: Electro cautory with
standard accessories quantity 10
standard accessories quantity 20
S. No.15 EQ2015: ELECTRO SURGICAL
Deleted.
CAUTERY UNIT
S.No. 64 EQ2064: Fully Automated
S.No. 64 EQ2064: Fully Automated
Chemiluminescence quantity 10
Chemiluminescence quantity 20
S.No. 65 EQ2065: Fully Automated
Deleted.
Chemiluminescence quantity 10
S.No. 79 EQ2079: Plate Centrifuge quantity 20
S.No. 79 EQ2079: Plate Centrifuge quantity 30
S.No. 94 EQ2094: Plate Centrifuge
Deleted.
Annexure-II (Specification)
S.No. 1 EQ2001: 5 Part Differential Hematology Analyzer
It should give 33 or more parameters i.e. WBC,
It should give 30 or more parameters i.e. WBC,
RBC, HGB, HCT, MCV, MCH, MCHC, CHCM,
RBC, HGB, HCT, MCV, MCH, MCHC,
RDW, HDW, CH, CHDW, PLT, MPV,PDW,
CHCM, RDW, HDW, CH, CHDW, PLT,
PCT, % RETIC, # RETIC , MCVr, CHCMr,
MPV,PDW, PCT, % RETIC, # RETIC , MCVr,
RDWr, HDWr, CHr, CHDWr., Asolute & %
CHCMr, RDWr, HDWr, CHr, CHDWr.,
values for NEUT, LYMPH, MONO,EOS, BASO, Asolute & % values for NEUT, LYMPH,
LUC, morphology results (user definable) like
MONO,EOS, BASO, LUC, morphology results
WBC : Left Shift , Atypical Lymph , Immature
(user definable) like WBC : Left Shift , Atypical
Granulocytes RBC: NRBC, ANISO, MICRO,
Lymph , Immature Granulocytes RBC: NRBC,
MACRO, RBC Ghosts.
ANISO, MICRO, MACRO, RBC Ghosts.
It should have high throughput of 100 or more
It should have high throughput of 100 or more
c.
d.
e.
f.
g.
h.
i.
j.
4.2
a.
b.
c.
d.
4.3
a.
b.
samples per hour in CBC and CBC / Diff. mode &
74 samples per hour or more in Retics mode.
Dual differential count for WBC
Platelets – Should have Dual angle Light Scatter
RBC – Should have light Scatter
HGB – Should have photometric and direct
cellular measurement
Retics – Should have on board, light scatter
(fluorescent dye) for Reticulocytes.
It should had have extensive linearity as
WBC - 0.02- 400 x 103 /ul
RBC - 0 - 7.0 x 106 / ul
PLT - 5 - 3500 x 103 / ul
HGB - 0- 22.5 g/dl
RETIC – 0.2- 24.5%
It should be free of tubings & pinch valves
ensuring low maintenance requirement.
Sample volume required in all modes not to
exceed 175ul.
Dead volume required < 300 ul.
It should be capable of reporting unique additional
parameters such as CHr- reticulocytes HGB
measurement to aid in anemia treatment.
samples per hour in CBC and CBC / Diff. mode
& 60 samples per hour or more in Retics mode.
Dual differential count for WBC
Platelets – Should have Dual angle Light Scatter
or equivalent.
RBC – Should have light Scatter or equivalent.
HGB – Should have photometric and direct
cellular measurement or equivalent.
Retics – Should have on board, light scatter
(fluorescent dye) for Reticulocytes or
equivalent.
It should had have extensive linearity as
WBC - 0.02- 400 x 103 /ul
RBC - 0 - 7.0 x 106 / ul
PLT - 5 - 3000 x 103 / ul
HGB - 0- 22.5 g/dl
RETIC – 0.2- 24.5%
Deleted.
Sample volume required in all modes not to
exceed 175ul to 300 ul.
It should be capable of reporting unique
additional parameters such as CHr or
equivalent reticulocytes HGB measurement to
aid in anemia treatment.
Deleted.
Deleted.
Company should give 1 year warranty on machine.
Company should give the CMC Rates for 5 years
with price bid. CMC will start after expiry of
warranty period.
Company should have at least 3 installations of
Company should have at least one service
machine at Raipur and one service engineer
engineer posted at Raipur. Company will
posted at Raipur. Company will provide the
provide the complete address of service
complete address of service engineer for physical
engineer for physical verification.
verification.
S.No. 2 EQ2002: Fully Automated Chemiluminescence’s Immunoassay Analyzer
Point 2. The throughput of the system should be
Point 2. The throughput of the system should be
more then 200 test per hour.
100 or more test per hour.
Point 4. The sample capacity should be more than
Point 4. The sample capacity should be
150 samples with continuous loading of reagents
minimum 150 samples with continuous loading
and samples.
of reagents and samples.
Point 9. System must clean the reagent probe and
Deleted.
sample probe automatically for all the parameters.
No manual cleaning of probe is acceptable.
Point 10. Multivalent calibrators for most
Point 10. Calibrators for most common assays.
common assays.
S.No. 3 EQ2003: Fully Automated Clinical Chemistry Analyzer
Point 1. Fully automatic discrete, random access
Point 1. Fully automatic discrete, random access
analyzer for routine chemistries, drug assays,
analyzer for routine chemistries, drug assays,
turbidimetry & electrolytes etc. It should have
turbidimetry & electrolytes etc. It should have
atleast 8-10 open parameters.
open parameters.
Point 3. Should have a through put of at least 600
Point 3. Should have a through put of at least
tests per hour without ISE and 900 tests/hr with
400-600 photometric tests per hour and 200ISE
300 ISE tests per hour.
c.
d.
e.
f.
g.
h.
i.
j.
k.
4.4
a.
b.
4.5
a.
4.6
a.
b.
4.7
a.
4.8
a.
Point 7. Minimum reagent volume-150-450 μl.
Point 11. Diffraction grating wavelength from
295-700 nm or more
Point 13. Fluids, which can be analyzed-Serum,
plasma, urine and CSF
Point 15. Sample capacity: unlimited thru
continuous loading without interrupting the run
Point 20. Should be able to perform
monochromatic, end point assays, rate assays,
fixed point assays, turbidimetric assays
Point 23. For data management: Should have
Pentium based PC touch screen monitor and
alphanumeric key board, should be LIS
compatible, bidirectional interfacing should be
possible. Laser Printer should be provided.
Point 26. The system should be FDA approved.
Point 28. Company should have at least 3
installations in Raipur and one service engineer
posted at Raipur. Company should give the
complete postal address of service engineer for
physical verification.
Point 29. Company should give one year warranty
on machine and warranty will start after successful
installation of machine.
S.No. 4 EQ2004: TonoRef II Autorefractor/
Keratometer, Non-Contact Tonometer
combined autorefractor, autokeratometer, and
Non-Contact Tonometer functionality bundled into
one compact, easy to use instrument.
Point 7. Minimum reagent volume-90-450 μl.
Point 11. Diffraction grating / filters
wavelength from 340-700 nm or more
Point 13. Fluids, which can be analyzed-Serum,
plasma, urine and CSF for whole blood.
Point 15. 150 or more sample loading
capacity with continuous loading.
Point 20. Should be able to perform
monochromatic, bichromatic, end point assays,
rate assays, fixed point assays, turbidimetric
assays
Point 23. For data management: Should have
suitable PC touch screen monitor and
alphanumeric key board, should be LIS
compatible, bidirectional interfacing should be
possible. Laser Printer should be provided.
Point 26. The system should be CE or USFDA
approved.
Point 28. Company should have at least one
service engineer posted at Raipur. Company
should give the complete postal address of
service engineer for physical verification.
Deleted.
S.No. 4 EQ2004: Autorefractor/ Keratometer,
Non-Contact Tonometer
combined autorefractor, autokeratometer, to be
supplied with automatic Non-Contact
Tonometer functionality bundled into one
compact, easy to use instrument.
S.No. 5 EQ2005: Ophthalmic Refraction Chair Unit
These full motorized recline units facilitate up and These full motorized recline units facilitate up
down movements for retinal examination lens
and down movements for examination with
with illuminating soft light for examination
illuminating soft light for examination
S.No. 6 EQ2006: Slit Lamp
Weight
8.10 kg (without power supply, Deleted.
headrest and options)
Add: Should be provided with motorized table.
S.No. 8 EQ2008: Rhino pharyngo Laryngo Fiberscope with accessories Case Pressure
compensation cap Lenkage tesfer
Office model, direction of view o, opening angle
Office model, direction of view o, opening
70, disfal tip outer diameter 3.5mm working length angle 70-80, disfal tip outer diameter 3.3 to
30 cm for connection to fibre optic length cable
3.5mm working length 30 cm for connection to
(495NL/NW) with Case Pressure compensation
fibre optic length cable (495NL/NW) with Case
capr for verification during gas sbcsilization
Pressure compensation capr for verification
Leakage tester w ith bulp & manometers
during gas sbcsilization Leakage tester w ith
bulp & manometers
S.No. 11 EQ2011: Electro cautory with standard accessories
Max RF power 400 watt ,Micro controller based,
Power 300-400 watt ,Micro controller based,
Tissue response circuit present, Patients plate
Tissue response circuit present, Patients plate
moniter present .Cut model Pure cut Blend 1,
moniter present .Cut model Pure cut Blend 1,
Blend 2, Endo cut, Coag mode, Spray coag,
Blend 2, Endo cut, Coag mode, Spray coag,
Dessicate, Fulgurate, Bipolar mode Micro,
MacroCut
4.9
a.
b.
c.
d.
e.
f.
4.10
a.
b.
4.11
a.
4.12
a.
b.
c.
d.
e.
4.13
a.
S.No. 17 EQ2017: Blood gas analyzer
Point 2. Measured parameters: pH, pCO2, pO2,
tHb, SO2, Na+, K+, Ca++, Cl-, Glucose, BUN
Point 7. FDA approved machine.
Point 11. Cassette storage must be at room
temperature expect metabolite cassettes
Point 12. True gas calibration
Point 13. Machine should be able to analyze the
sample within 2-3 minutes after switching it ‘on’
Point 14. Should be supplied with starter kit
reagents and quality control material for at least 1
year use.
Dessicate, Fulgurate, Bipolar mode Micro,
MacroCut
Add: LCD or LED display, Irrigation, Visual
and audio alarms, min. 10 memory settings,
bipolar coagulation with auto start facility,
monopolar cut and coagulation, disposable
cautery pencils (10 nos.), autoclavable cords.
Add: Product should be IEC, USFDA
approved.
Point 2. Measured parameters: pH, pCO2, pO2,
tHb, SO2, Na+, K+, Ca++, Cl-, Glucose,
Bilirubin, Lactate & Coximetry.
Point 7. CE or USFDA approved machine.
Deleted.
Point 12. True gas calibration / Liquid
Calibration.
Deleted.
Point 14. Should be supplied with starter kit
reagents for 500 tests and quality control
material for at least 1 year use which should be
run after every 350 tests.
S.No. 19 EQ2019: Arterial Blood Gas Analyzer- Cartridge based
Use life: 28 Days Number 250, 400of tests:750'
Use life: 60 Days Number 250, 400of tests:750'
Add: Should measure parameters: pH, pCO2,
pO2, tHb, SO2, Na+, K+, Ca++, Cl-, Glucose,
Lactate and calculated parameters.
S.No. 22 EQ2022: Deep Freezer
Add: Capacity- 300ltr and above
S.No. 26 EQ2026: Atomic Absorption Spectrophotometer (Graphite Model )
Point 1. PC based high performance; Optical
Point 1. PC based high performance; True
Double Beam Atomic Absorption
Double Beam Atomic Absorption
Spectrophotometer operating on WINDOWS Vista Spectrophotometer operating on WINDOWS 8
based software with 02 ton Air conditoner
based licensed software with 02 ton Air
conditoner
Point 4. New 3D double beam optics, optimal light Point 4. Double beam optics, optimal light beam
beam alignment system and digital light filter for
alignment system and digital light filter for high
high sensitivity.
sensitivity.
Point 12. Dual background correction as standard - Point 12. Dual background correction as
High speed dual frequency simultaneous
standard - High speed dual frequency
measurement D2 background correction and
simultaneous measurement D2 background
correction
preferably high-speed Self Reversal
background correction
Point 18. Software operation on Windows Vista
Point 18. Software operation on Windows 8
OS for complete instrument control, data
licensed OS for complete instrument control,
acquisition and processing
data acquisition and processing
Point 21. The system should have all the Hollow
Point 21. The system should have all the Hollow
Cathode lamps for the required analysis.
Cathode lamps for the required analysis i.e., 8
nos. including Ca, Zn, Mg, Cu, Fe, Li, Co,
Selinium.
S.No. 27 EQ2027: Ultra Centrifuge
Point 16. On-board rotor library includes
Point 16. On-board rotor library includes rotors
b.
c.
4.14
a.
b.
4.15
a.
4.16
a.
Bookman Coulter rotors and tube configuration.
Point 17. Air conditione 02 ton capicity
and tube configuration.
Deleted.
Add: One fixed angle rotor capable of going up
to 1,00,000RPM (about 8,02,000g,
volume per tube 6 ml approx) along with
necessary tubes and accessories.
One swing bucket rotor capable of
running up to 32,000RPM (about
1,75,000g, volume per tube 38ml approx)
along with tubes including conical tubes
and adapters and accessories.
Adaptors to run volumes of 33.0ml,
27.0ml, 15.0ml and 8.5ml volumes using
Conical tubes for up to 1,75,000g.
Necessary Voltage stabilizer should be
provided.
S.No. 31 EQ2031: High Defination DH Camera control unit with 3 chip camera with 23'' Flat
screen monitor system with 0° 30°, 45°, 70° Telescope
Autoclavable Hopkins Telescope Diameter 4mm
Autoclavable Full HD Hopkins Telescope
& 2.7mm Length 18cm,0°, 30°, 45°, 70°
Diameter 4mm & 2.7mm Length 18cm,0°, 30°,
Telescope HD Camera control unit for use with
45°, 70° Telescope HD Camera control unit for
HD 3 chip camera head, max. resolution 1920 x
use with Full HD camera head, max. resolution
1080 pixel, with integrated digital image, Power
1920 x 1080 pixel, with integrated digital
supply 100-240 v. 50/60 Hz, including main cord 2 image, Power supply 100-240 v. 50/60 Hz,
X connecting cable
including main cord 2 X connecting cable
DVI - connecting cable, SCB - connecting cable,
DVI - connecting cable, SCB - connecting
Keyboard with US english character set,
cable, Keyboard with US english character set,
3 chip full HD Camera head autoclavable,
full HD Camera head sterilizable, resolution
resolution 1920 x 1080 pixels, Progressive scan,
1920 x 1080 pixels, Progressive scan, soakable
soakable gas and plasma sterilization with
gas and plasma sterilization with integrated
integrated parfocal zoom lens, Focal length F=15- parfocal zoom lens, Focal length F=15-31mm, 2
31mm, 2 freely programmable camera head
freely programmable camera head buttons for
buttons for use with colour system PAL/NTSL
use with colour system PAL/NTSL
HD - Monitor 23''
26 inch LED medical grade monitor
Fibre optic light cable
Fibre optic light cable
Cold light countain halogen
LED Light Source- 175 watt or more
Power supply 100/120/230, 240 VAC
Power supply 100/120/230, 240 VAC
50/60 Hz, including main cord
50/60 Hz, including main cord
Add: Should be USFDA approved.
S.No. 32 EQ2032: 4ft, Class II, Type A2 Biosafety Cabinet
99.99% efficient HEPA filter, Filter remaining
99.99% efficient HEPA filter, Filter remaining
graph, 230V 50 Hzno speed variation due to
graph should be displayed on LCD, 230V 50
voltage fluctuation, base stand for cabinate, UV
Hzno speed variation due to voltage fluctuation,
light, inflow velocity not less than 100fpm,
base stand for cabinate, UV light, inflow
downflow velocity not less than 50fpm, 70%
velocity not less than 100fpm, downflow
recirculation.
velocity not less than 50fpm, 70%
recirculation.
Add: Should be CE, NSF/EN certified.
Should have Front made up of Tuffon
Glass.
S.No. 33 EQ2033: ELISA reader with washer
Fully Automated and programmable with
Plate reader should have following features
computer
1. Detection Mode: Absorbance
2. Light Source: Log life LED, self-calibrating
3. Photo detector: silicon Photo Diode
4. Plate type: 96 well microplate including
strip well microplate
5. Measurement Mode: Endpoint, kinetic and
multi- label measurements
6. Measurement Channel: 8 Channel
7. 4 Different Shaking mode
8. Measurement range:0-4 OD
9. Resolution : 0.0001 OD or better
10. Wavelength Range: 400-750nm
11. Wavelength selection up to 8 filters
12. Standard filters 405nm, 450nm, 492nm,
620nm
13. Accuracy 0-2OD <(0.5%+0.005OD) or
better
14. Precision : ≤(0.5% +0.005OD) or better
15. Linearity : 0-2 OD : ≤ 1% or better
16. Fully qualitative and quantitative EIA
analysis including all major curve fittings
and comprehensive QC function.
17. Window based software
18. Display: Operation through PC and Printer
Plate Washer should have following features
1. Cross wise aspiration, over flow washing,
bottom washing
2. Automatic manifold detection
3. 8 or 16 channel manifold
4. Manifold, plate carrier auto cleavable and
also can be sterilized
5. 1-4 wash liquid channel option- Standard -2
6. Equipment should be un-pressurized,
capable of using any bottle or container.
7. Cell washing should be standard feature
8. Volume of wash liquid dispensed
9. 50 to 3000µl with 50 1 increment
10. Suitable for U, V & flat bottom micro plate
etc.
11. Residual volume per well <2µl
12. Up gradation for magnetic bead and
vacuum filtration
Both of system should be FDA, IVD, CE and
ISO certified.
4.17
a.
S.No. 34 EQ2034: PCR work station
Capable of performing PCR
1. System should be open, modular platform
& able to performed Nucleic Acid
extraction & PCR cleanup etc.
2. System should be capable of performing
Nucleic Acid extraction using vacuum as
well as magnetic bead based technique.
3. System should have at least 9 Deck position
& 18 MTP storage position or more.
4. System should be equipped with single
channel pipetting arm with flow through
technique to perform any pipetting ranging
from 0.5 ul to 5ml. Arm should have
following features.
 Independent Z & Y axis control
 Contact Liquid Level Sensing
 Flexibility of standard washable Tips &
optional Disposable Tips
 Integrated wash station to clean the
needles internally and externally
 Disposable Tip sensing features.
 Pipetting precision for Volume 10 ul
<3.5% , Volume 100ul <1% or better.
 Pipetting arm capable of handling
different liquids with different physical
properties like density, viscosity and
vapor pressure
 Should be able to handle test tubes, vials
tubes of different size and 96-384 well
plates.
5. System should be offer with carrier for
carrying 96 test tubes, & MTP 96 3 position
carrier as per application need.
6. System should be offer with magnetic bead
module.
7. System should be offer with programming
software & computer i5 Processor PC,
500GB HDD, 4 GB RAM, 22’’ TFT.
8. System should have validated protocol for
commercial available extraction kits.
Optional Up gradation in future
1. System should have flexibility to upgrade
additional independent Arm for plate
transport, on Deck & Off Deck integration.
2. System should have flexibility to upgrade
system from single channel to 8 channel
system pipetting system.
3. Vacuum module with 2 independent
position to perform sextraction
4. Temperature controlled carrier on deck to
perform PCR assay.
4.18
a.
S.No. 35 EQ2035: Electronic balance
Able weigh from 0.001 to 10 gm, accuracy ±
0.0003
b.
4.19
S.No. 38 EQ2038: Operating Table
Add: Made up of 304 SS & suitable for
ophthalmology department.
a.
4.20
S.No. 39 EQ2039: B Scan
S.No. 42 EQ2042: Fundus Camera (Zeiss)
Add: Should be supplied with printer, UBM
system
S.No. 42 EQ2042: Fundus Camera
S.No. 43 EQ2043: Phacoemulsification machine
Paristaltic pump (PPS) programmbale double
Paristaltic / venturi pump (PPS) programmbale
a.
4.21
a.
4.22
a.
Able weigh from 0.001 to 60 gm, accuracy ±
0.0001 gm
Add: Should be supplied with standard
accessories like:
1. Certified & calibrated weights (1mg, 10mg,
50mg, 100mg & 500mg)
2. Specific gravity measurement kit.
4.23
a.
4.24
a.
b.
c.
d.
e.
4.25
a.
b.
c.
d.
e.
f.
g.
4.26
a.
linear foot pedal single multiple and continuous
double linear foot pedal single multiple and
burst modes vacuum rise time up to 25 levesl pulse continuous burst modes micro pulse
rat from 1 to 40 Hz big LCD screen
Hand piece: 4-6 crystals, 28 to 40 kHz
Vacuum: 0 to 500 mmHg
Diathermy, Electrical / Pneumatic vitrectomy
cutter speed upto 0-1500 cuts/sec
Titanium Phaco tip- 2.8mm tip
With all standard standard accessories
S.No. 45 EQ2045: High Definition (HD) Camera control Unit with 3 chip camera with 23" flat
Screen Monitor System with 0,30,45,70 Telescope
Add: LED Light Source
S.No. 49 EQ2049: Endoscopic Sinus Surgery
S.No. 49 EQ2049: Upper GI Endoscope
Instrument
(Gastroscope)
Magnification :
Magnification :
 Should be atleast 8X with atleast 2X elctronic  Should be atleast 8X with atleast 2X
zoom
elctronic zoom / Full HD Processor
Technical Specification :
Technical Specification :
 Depth Of Field – Normal – 5- 100 mm ; 6 Depth Of Field – Normal – 5- 100 mm ; 6100 mm. Near 2- 6 mm ; 2- 3 mm
100 mm. Near 2- 6 mm ; 2- 3 mm / close
focus
 Add: Should be fitted with megapixel CCD
for Full HD image
LIGTH SOURCE :LIGHT SOURCE : Examination Lamp – Xenon lamp atleast 300  Suitable LED Light Source should be
watts and 2 extra lamps to be provided
provided.
 Average lamp life – Atleast 500 hours
 Emergency lamp – Halogen 35 to 75 watts
 Should have integrated Air and Water supply
pump with minimum 3 output levels.
Add: Should be provided with standard
accessories like Biopsy Forcep, Balloon Dilator,
Guide Wire, etc.
S.No. 50 EQ2050: Fully Autoanalyzer
Point 1. System: Fully automated, random access,
Point 1. System: Fully automated, random
floor model biochemistry analyzer with a
access, floor model biochemistry analyzer with
minimum throughput of 800tests/hour, 400tests
a minimum throughput of 800tests/hour, 500photometry/hr, 400ISE/hr
600 tests photometry/hr, 200-300 ISE/hr
Point 2a. Accommodate a minimum of 76 reagent Point 2a. Accommodate 40 or more reagent on
on board.
board.
Point 2f. Ability to programs & use up to 120 User Point 2f. Ability to programs & use up to more
Defined Reagents (third party reagents) defined at than 100 User Defined Reagents (third party
any given time
reagents) defined at any given time
Point 3a. Continuous loading of samples through
Point 3a. Continuous loading of samples
sample racks (minimum capacity to handle 100
(minimum capacity to handle 40 or more
samples at a time)
samples at a time)
Point 3c. Cooled positions for handling STAT
Deleted.
samples, Calibrators & quality control materials
Add: 2 reagent probes and smple probesshould
be provided.
Point 4b. Cooled positions for Quality control
Point 4b. Reserved position to run quality
material
control separately.
S.No. 51 EQ2051: Semi Autoanalyzers
OPERATOR INTERFACE:
OPERATOR INTERFACE:
Membrane keyboard, for direct function and
Membrane keyboard, for direct function and
b.
c.
4.27
a.
alpha-numeric entry
alpha-numeric entry / Touch screen
MULTIPLE TESTING:
MULTIPLE TESTING:
Up to nine replicates.
Up to three replicates.
DELAY TIME:
DELAY TIME:
programmable, 0 to 000 seconds
programmable, 0 to 999 seconds
S.No. 53 EQ2053: Automated Elisa With reader and washer
Fully automated system
ELISA Reader System Specifications:
1. Analyzer type: 1 Plate ELISA reader,
throughput 1 plate/run
2. Reaction unit/Incubation capacity : 1 plate
3. Mixing capability: Yes
4. Reagent System Open
5. Reaction / reading system:96 well
microtiterplate
6. Tests Measuring modes EIA photometric
7. Analysis modes Endpoint
8. Calculation algorithms ABS, Single
Standard, point to point, kinetic, %ABS,
liner regression, exponent regression,
logarithm regression, power regression,
cubic spline, 4 PL
9. Number of programmable tests Unlimited,
number
of
programmable profiles
Unlimited
10. Reading Optical system Filter wheel,
continuously rotating
11. Reading Single or double- wavelength
simultaneous reading
12. Light source: Halogen 6.5 V / 19W or 8V
/50W
13. Spectral range: 400-700 nm
14. Wavelengths
filters
preinstalled:405,450,492,630 nm
15. Dynamic range (measurement range):0-3.5
A(0-2, 0 A)
16. Accuracy:0-2.0 A : ± 1.0 % or ±0.007 A
17. Precision 0-2.0 A : ± 0.5 % or ± 0.005 A
18. Detector Photodiode detection
19. Linearity: 0-2.0 A ± 2.0% or ± 0.007 A
20. Data Processing interfaces: RS-232C serial
interface, 2x USB interface, 1x SD card
slot
21. LIS/LIMS: Prepared
22. Multiple assays per plate 1assay per plate
23. Memory capacity 1000 patients and 10000
sample record
24. Number of programmable tests 200 test
25. Touch panel and pen, external mouse
26. Supported operating system Windows
27. CE Marked
28. External printer enabled
Automatic ELISA plate Washer, Shaker and
Incubator:
1. Analyzer type: 1 plate washer & shaker, 2
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
4.28
a.
4.29
a.
4.30
a.
4.31
a.
4.32
a.
b.
4.33
a.
b.
c.
d.
e.
plate incubator
Throughput (Max): 1 plate run wash &
shake, up to 2 plates incubation
Reaction / reading system 96 well
Microtiterplate
Operation mode batch
Reaction Unit/ Incubation Capacity 1
chamber for up to 2 plates
Mixing capability: Mixing in wash
position
Incubation temperature 25°/30°/37.0°C
fixed
Incubation time free choose able
(increment 1 minute)
Washing capacity 3x wash buffer bottles
2L
Wash head 8 channels (included)
Plate type Flat, round and V-Bottom wells
Wash mode plate and strip mode
Dispense volume 10-3000µl/well in 1µl
steps
Features 10 wash profiles
Up to 99 wash cycles
Data processing interfaces RS 232
Number of programmable tests: 100 wash
protocols
User interface LCD Display
18.
S.No. 54 EQ2054: Automated Blood culture system
Fluoroscent base, 50 cultures simultaneously,
Fluoroscent / colorimetric / air pressure base,
alarm for growth
50 cultures simultaneously, alarm for growth
S.No. 56 EQ2056: Electronic weighing Machine
Able to weigh 0.5- 20 grams
Least Count- 1 mg
Capacity- 500 to 600 gm
S.No. 57 EQ2057: Deep Fridge
Add: Capacity- 300 ltr.
S.No. 60 EQ2060: Spectrophotometer
UV Visibledouble beam with multiple cuvette
UV Visibledouble beam with multiple cuvette
holders with attached PC, inbuilt stabilizer, Two
holders with attached PC, stabilizer, Two sets of
sets of quartz cuvette, digital display with software quartz cuvette, digital display with software
control system.
control system.
S.No. 62 EQ2062: Ultra centrifuge
Point 3. Speed control accuracy of +/- 1 RPM and Point 3. Speed control accuracy of +/- 10 RPM
temperature control of +/- 0.5°C.
and temperature control of +/- 0.5°C.
Point 16. On-board rotor library includes
Point 16. On-board rotor library includes rotors
Beckman Coulter rotors and tube configurations. and tube configurations.
S.No. 64 EQ2064: Fully Automated Chemiluminescence
Point 1. System should be based on technology of
Point 1. System should be based on technology
paramagnetic particles with dio oxetebce
of Chemi-Luminescence.
phosphtas.
Point 2. At least 22 assays from one sample.
Point 2. At least 18 assays from one sample.
Point 4. Loading of minimum 50 samples at one
Point 4. Loading of minimum 30 samples at one
time
time
Point 6. Minimum 100 tests/hour throughput
Point 6. Minimum 60 tests/hour throughput
Point 10. Self-sealing reagent packs, pack size 2 x Point 10. Self-sealing reagent packs, pack size
f.
g.
4.34
a.
b.
c.
d.
e.
f.
4.35
a.
50 tests for all parameters.
Point 12. Any no. Of STAT samples can be
loaded at given point of time.
Point 20. Should cover test menu for Anemia,
Reproductive, Thyroid panel, Cardiovascular,
Skeletal, Tumor marker panels, etc…,
more than 50 tests for all parameters.
Point 12. Any sample can be loaded as STAT at
given point of time.
Point 20. Should cover test menu for Anemia,
Reproductive, Thyroid panel, Cardiovascular,
Skeletal, Tumor marker panels, etc…,
Minimum no. of tests- 60
S.No. 66 EQ2066: Atomic Absorption Spectrophotometer (Flame Furance
Fully controlled by PC, Automatic wavelength
Fully controlled by PC, Automatic wavelength
Set-up, Automatic Slit switching, Automatic Lamp Set-up, Automatic Slit switching, Automatic
current and gain set-up within 40 seconds. Auto
Lamp current and gain set-up within 40
switching and control 6 lamps on rotating turret,
seconds. Auto switching and control more than
Automatic flame ignition, automatic flow control
6 lamps on rotating turret, Automatic flame
for gases and gas leak detection, nitrous oxide /
ignition, automatic flow control for gases and
acetylene gas selection (optional) to analyze
gas leak detection, nitrous oxide / acetylene gas
elements under the condition of high temperature,
selection to analyze elements under the
the use of advanced grating for strong energy and
condition of high temperature, the use of
high resolution., compatible with MS windows XP advanced grating for strong energy and high
and user-friendly graphical user interface.
resolution., compatible with MS windows XP
and user-friendly graphical user interface.
Add: Integrated Flame & Furnace True Double
Beam System with suitable latest Windows
based software.
Print Output :- Calibration Curve , Signal
Print Output :- Calibration Curve , Signal
Spectrogram , Instrument Conditions , Analysis
Spectrogram , Instrument Conditions , Analysis
Parameters, Analysis Results and others all can be Parameters, Analysis Results and others all can
stored automatically and printed. Auto sampler for be stored automatically and printed. Auto
Flame Up to 60 ea of 5ml vials, carousel type
sampler for Flame or furnace Up to 60 each of
5ml vials
(OPTIONAL)
Description of goods/Qty: Point 11.
Deleted.
Installation charges for fume hood Rs. 2000/- per
feet extra if required above 10 feet. – 1 no.
Recommended Accessories: Point 2.
Recommended Accessories: Point 2.
HP Deskjet Printer – 01 no.
Printer – 01 no.
Recommended Accessories: Point 2.
Recommended Accessories: Point 2.
HP Laserjet Color Printer (HP2600) – 01 no.
Color Printer – 01 no.
S.No. 68 EQ2068: Real Time Polymerase Chain Reaction Equipment
96-well (not
96-well (not
interchangea
interchangea
Thermal
Thermal
ble)
ble)
Block
Block
5- 50 µl
Block
(PikoReal 24,
10-50 µl
Formats:
5-20 µl (Piko
Block
Consumable 96-well PCR
Real 96)
Formats:
Plate
s
24-well or
Max Heating
Consumable 96-well Piko
≥4.5°C/sec
rate
PCR Plate
s
Max cooling
Max Heating
rate
4.5°C/sec
>5°C/sec
rate
Temperature
Max cooling
range
4- 99.9°C
rate
4.5°C/sec
Temperature
Temperature 4- 99.9°C
accuracy
±0.2°C
range
Temperature
accuracy
Temperature
uniformity
Heated lid
Temperature
range
Control
Optics
Excitation
Excitation
range
Detection
Detection
range
Detection
channels:
Exitation
(nm)
Dyes
475 - 500
Green
515 – 535
Yellow
570 – 590
600 – 640
475 – 500
HRM,
Temperature
uniformity
Heated lid
Temperature
range
±0.2°C
±0.3°C at
95°C
30 - 110°C
Automatic
Temperature
and Pressure
setting
5 LEDs
475 - 640 nm
CCD
520 - 740 nm
Emission
(nm)
520 - 550
557 - 590
615 - 650
666 - 740
520 - 590
Pre Calibrated
FAM, SYBR
HEX,
Yakima
ROX, Texas
Red
Cy 5
SYBR Green
for
Control
Optics
Excitation
Excitation
range
Detection
Detection
range
Detection
channels:
Exitation
(nm)
Dyes
475 - 500
Green
515 – 535
Yellow
±0.3°C at
throughout
applicable
range
30 - 110°C
Automatic
Temperature
and Pressure
setting
5 LEDs
475 - 640 nm
CCD
520 - 740 nm
Emission
(nm)
Pre Calibrated
520 - 550
FAM, SYBR
557 - 590
HEX,
Yakima
570 – 590
600 – 640
615 - 650
666 - 740
475 – 500
HRM,
520 - 590
Single plex
FAM,
Single plex
FAM,
Multiplex
HRM
Dynamic
range
Sensitivity
Scan time for
SYBR Green
Up to 4
targets
Available
2011
10 orders of
magnitude
1 copy
<10Sec
ROX, Texas
Red
Cy 5
SYBR Green
for
Multiplex
HRM
Dynamic
range
Sensitivity
Scan time for
4
SYBR Green
Up to 4
targets
Available
2011
10 orders of
magnitude
1 copy
<10Sec
4
multiplexing
channels
Software
Analysis
modes
Absolute
quantification
Relative
quantification
Melt curve
analysis
Allelic
discriminati
on
System
Operating
systems
Communicat
ion
single pc
Windows
XP,
Windows 7
Ethernet, up
to 10
instruments
can be
operated
200W
maximum
300 x 230 x
310 mm
10 kg'
4.36
a.
b.
c.
d.
e.
4.37
a.
b.
c.
Power usage
Dimensions
(W x D x H)
Weight
2KVA 3hrs
UPS
S.No. 69 EQ2069: Fully Autoanalyzer
Sample:- Point 1. Sample positions 60
Sample:- Point 1. Sample volume (1.0-100µl) in
increments of 1µl
Reaction Module:- Point 2. Reaction temperature
37±0.1 ( dry air incubation).
Software:- Point 1. Windows 2000/XP platform.
Dimensions:1. 130cms X 100cms X 85cms.
2. 90 Kg.
S.No. 74 EQ2074: Auto refracto keratometare
Sphere (SPH): -30 to+22D when VD = 12mm
(increments:0.12D and 0.25D)
Corneal Astigmatism: 0 to – 15 D
Operating Range
Front/back: 40mm
Lent/right: 90mm
Up / down: 30mm
Dimensions (W x L x H) Approx. 260 x490 x
multiplexing
channels
Software
Analysis
modes
System
Operating
systems
Communicat
ion
single pc
Power usage
2KVA 3hrs
UPS
Absolute
quantification
Relative
quantification
Melt curve
analysis
Allelic
discriminati
on
Windows 8,
Windows 7
Ethernet, up
to 10
instruments
can be
operated
200W
maximum
Sample:- Point 1. Sample positions 25-30
Sample:- Point 1. Sample volume (3-100µl) in
increments of 1µl
Reaction Module:- Point 2. Reaction
temperature 37±0.1
Software:- Point 1. Windows 7/8 platform.
Deleted.
Sphere (SPH): ±25D when VD = 12mm
(increments:0.12D and 0.25D)
Corneal Astigmatism: 0 to ±10D
Operating Range
Front/back: 40mm or more
Left/right: 90mm or more
Up / down: 30mm or more
4.38
a.
b.
470mm
S.No. 75 EQ2075: Analytical Electronic Balance
Electronic analytical balance with built –in,
motorized weight, dynamic temperature
compensation, Backlit (blue color ) 8 segment
display, 2 large tare keys, stability indicator, 4 user
– selectable filter level, plus optimized display
update rate.
Pan (inner diameter)* :80mm
Weighting Chamber (WxDxH):174x181x240 mm
Balance housing(WxDxH) : 213x342x340 mm
Net weight : approx 6.7 kg
*) triangular weighing pan shape
∅ = diameter of the inner circle
c.
4.39
a.
4.40
a.
4.41
a.
4.42
a.
b.
4.43
a.
b.
4.44
a.
b.
S.No. 76 EQ2076: Blood Auto Analyzer Machine
Electronic analytical balance with built –in ,
motorized weight, dynamic temperature
compensation, Backlit 8 segment display, 2
large tare keys, stability indicator, 4 user –
selectable filter level, plus optimized display
update rate.
Deleted.
Add: Should be supplied with standard
accessories like:
1. Certified & calibrated weights (10mg,
50mg, 100mg, 500mg & 100gm)
2. Specific gravity measurement kit.
S.No. 76 EQ2076: Automatic Blood Gas
Analyser
S.No. 79 EQ2079: Plate Centrifuge
Dimensions (H x W x D) 405 x 490 x 685 mm
Deleted.
Weight (without rotor) 82 kg
S.No. 84 EQ2084: Powered Electrophoresis migration System
Should have in built power supply.
Should be provided with power supply as
standard package.
S.No. 86 EQ2086: -35ºC Deep Freezer
Point 2. Capacity : 475 to 500 Ltr.
Point 2. Internal Capacity : 475 to 500 Ltr.
Point 12. Eye level digital display with both set
Point 12. Digital display with both set and
and actual temperature
actual temperature
S.No. 88 EQ2088: Micropipettes (Multi channel) Volume Range: 1 – 20 µL
1 µL to 20 µL and 20 µL to 300 µL.
1 µL to 20 µL.
SpecificationsSpecificationsVolume
Systematic
Random
Volume
Systematic
Random
(µL)
error(µL)
error(µL)
(µL)
error(µL)
error(µL)
1
±0.1
≤0.08
1
±0.1
≤0.08
2
±0.1
≤0.08
2
±0.1
≤0.08
10
±0.2
≤0.1
10
±0.2
≤0.1
20
±0.4
≤0.2
20
±0.4
≤0.2
20
±1.00
≤0.35
30
±1.00
≤0.45
150
±1.50
≤1.50
300
±3.00
≤3.00
S.No. 89 EQ2089: Micropipettes (Multi channel) Volume Range: 20 – 300 µl
1 µL to 20 µL and 20 µL to 300 µL.
20 µL to 300 µL.
SpecificationsSpecificationsVolume
Systematic
Random
Volume
Systematic
Random
(µL)
error(µL)
error(µL)
(µL)
error(µL)
error(µL)
20
±1.00
≤0.35
1
±0.1
≤0.08
30
±1.00
≤0.45
2
±0.1
≤0.08
150
±1.50
≤1.50
10
±0.2
≤0.1
4.45
a.
4.46
a.
4.47
a.
b.
c.
4.48
20
±0.4
≤0.2
20
±1.00
≤0.35
30
±1.00
≤0.45
150
±1.50
≤1.50
300
±3.00
S.No. 93 EQ2093: Deep Freezer (-80° C )
Technical Specifications:
 Capacity : 368 ltrs.
 Communication : Rs 232 Port (Optional)
S.No. 95 EQ2095: UV-VIS Spectro- photometer
 Dimensions (W×D×H) : 450(W) x 490(D) x
270(H)
 Weight : 15kg
S.No. 98 EQ2098: Laminar Air Flow Batch
Work Table : Wooden with melamine finish.
300
±3.00
≤3.00
Technical Specifications:
 Capacity : 700 ltrs.
 Communication : Rs 232 Port or Data
Logger
Deleted.
S.No. 98 EQ2098: Laminar Air Flow Bench
Work Table : Wooden with melamine finish /
Stainless Steel
Add: Filter Location should be horizontal.
S.No. 102 EQ2102: Multichannel Pipettes
Add: Capacity: 10-100 µl
30-300µl
a.
100-1000µl
No. of Channels: 8
Battery:- Lithium ion rechargeable, located in base Battery:- Lithium ion rechargeable, located in
b.
of pipette (not shown).
base of pipette.
Power Supply:- AC adapter for recharging battery Power Supply:- AC adapter for recharging
c.
battery.
(not shown).
4.49 S.No. 103 EQ2103: Automated blood culture system
Must accommodate minimum 100-120 vials at a
Must accommodate minimum 90-120 vials at a
a.
time.
time.
4.50 S.No. 104 EQ2104: Automated ID/AST Bystem
a.
Add: Capacity: More than 50 samples at a time.
All other terms and conditions of the tender enquiry remain unaltered.
NOTE: Bidders submitting any document in language other than Hindi / English then an English
translation of the same should be submitted in affidavit form (Stamp paper).
Managing Director
CGMSC Ltd., Raipur Chhattisgarh