PV .035 Digital IVUS Catheter Data Sheet

Radiopaque and inked catheter shaft
markers facilitate low contrast and low
radiation procedures
UTILIZE THE
POWER OF
IVUS TO HELP
REDUCE
RADIATION:
ADJUNCTIVE
IVUS GUIDED
EVAR/TEVAR
PROCEDURES
Thoracic Aorta Dissection
Renal Vein
and Left Renal Artery
External Iliac Artery
with Internal Iliac
(Hypogastric) Branch
Study data reported a reduction in
contrast and radiation.1
140
FEATURE BENEFIT
AMOUNT OF CONTRAST (CC)
120
„„
Confirms
IVUS IMAGING
position in true lumen
versus false lumen
„„
Provides real-time diameter
measurements
„„
Assesses vessel morphology
100
80
60
40
20
WITHOUT
IVUS
WITH
IVUS
(n=31)
(n=79)
0
(IVUS group 67 ± 34 ml vs non IVUS group 123 ± 50 ml; p<0.01)
RO markers spaced
1 cm apart allow for quick
length measurements and
may reduce overall procedural
contrast and radiation use.
„„
Pullback
the IVUS catheter and
count the external inked markers
on proximal shaft to facilitate
reduced contrast and radiation
length assessments.
* Imaging plane
1. Hoshina, et al. A Retrospective Study of Intravascular Ultrasound use in Patients Undergoing Endovascular Aneurysm Repair:
Its Usefulness and a Description of the Procedure. Eur J Vasc Endovasc Surg. 2010;40:559-563.
2. von Segesser, et al. Routine Use of Intravascular Ultrasound for Endovascular Aneurysm Repair: Angiography is not
Necessary. Eur J Vasc Endovasc Surg. 2002;23:537-542.).
All IVUS Images: Data on File.
Study data reported a reduction in
mean x-ray exposure time when utilizing
adjunctive IVUS vs. angio alone.2
25
140
120
100
80
60
40
20
WITHOUT
IVUS
(n=79)
0
(IVUS group 67 ± 34 ml vs non IVUS group 123 ± 50 ml; p<0.01)
WITH
IVUS
(n=31)
MEDIAN X-RAY EXPOSURE TIME
INKED MARKERS,
1 CM APART,
ON PROXIMAL
SHAFT
IVUS GUIDED EVAR/TEVAR PROCEDURES
MAY HELP REDUCE PROCEDURAL
CONTRAST USE AND RADIATION TIME
„„
Visible
AMOUNT OF CONTRAST (CC)
25 RADIOPAQUE
MARKERS,
1 CM APART,
ON DISTAL SHAFT
20
15
10
5
WITH
WITHOUT
IVUS
24MIN
(n=31)
0
range 9-65 min
IVUS
(p<0.05)
range 0-60 min
8(n=49)
MIN
Inked Markers
(non-radiopaque)
Radiopaque
Markers
1 mm 5 mm
1 cm
1 cm
1.5 mm
1 cm
6.5 mm
1 cm 1 cm
12 mm
Imaging Plane
1.5 mm
PIM Connector
Y-
Co
90 cm Working Length
nn
ec
tor
Guide Wire
Exit Port
≤ 0.038"
(0.97 mm)
30 cm GlyDx® Hydrophilic Coating
7F
25 RO Markers,
1 cm apart
Image up to
60 mm diameter
Tip O.D.
≤ 0.055”
8.2 F
Transducer
Ordering Information: 88901
Visions® PV.035 Digital IVUS Catheter
Physician Payment CPT Codes – Peripheral
Frequency
10 MHz
+37250
IVUS of initial non-coronary vessel
Maximum Guide Wire
0.038"
+37251
IVUS of additional non-coronary vessels
Platform
OTW
PIM Connector Length
100 cm
75945-26
Radiologic supervision, initial non-coronary vessel
Minimum Sheath Size
8.5F
75946-26
Radiologic supervision, additional non-coronary vessels
Volcano Corporation
Corporate Headquarters
3721 Valley Centre Drive, Suite 500
San Diego, CA 92130 USA
Phone: 800-228-4728
Fax: 858-720-0325
Volcano Europe BVBA/SPRL
EU Authorized Representative
Excelsiorlaan 41
B-1930 Zaventem Belgium
Phone: +32 2 679 1076
Fax: +32 2 679 1079
Volcano Japan
Japan Headquarters
Hamamatsucho Square 6F
1-30-5, Hamamatsucho,
Minato-ku, Tokyo 105-0013, Japan
Phone: +81-3-6430-9400
Fax: +81-3-6430-9401
For further information about Volcano and its products, please visit www.volcanocorp.com.
Volcano, the Volcano Logo, Visions, and GlyDx are registered trademarks of Volcano Corporation.
601-0500.20/003