Purpose
To describe the procedure to unpack the processing module Module
i2000/i2000SR
2 Troubleshooting
Materials Required
Standard Tool Kit
1.0 hr
Time
3 Parts Lists
4 Removal & Replaceme
5 Procedures
.
Action
Steps
6 Planned Maintenance
7 Pre-Site Interview & Ins
8 Installation & Setup
Remove Packing
1. Remove the packing sides and top. [1]
Verify Packed Items
1. Remove the assorted parts kit, wrench, and ramp plate. [2]
2. Verify that all pieces are present.
Remove Wood Strip
1. At the RAMP END location of the shipping platform, remove the two (2) screws affixing the strip of wood to the
c8000 Checklist
Unpack
SCC - Hardware & C
Install Accessories
Processing Module P
Installation of Pressu
RSH Installation
SCC - Software Insta
RSH Calibration
Robotic Positions Se
RSH Verification
P-159 PVT
SCC Installation Veri
Installation Documen
i2000SR Checklist
Unpack
SCC - Hardware & C
Processing Module P
RSH Installation
SCC - Software Insta
RSH Calibration
Processing Module C
shipping platform.
2. Remove the strip of wood. [3]
RSH Verification
Processing Module V
SCC Installation Veri
Installation Documen
ci8200 Checklist
i2000 LAS Checklist
i2000 SSH Checklist
i4000 SSH Checklist
Upgrade
Relocation
Remove Hold-Down
Brackets
http://gssweb.add.abbott.com/architect/3a_i2000sr_html/3a_start.htm[5/14/2014 9:48:54 AM]
1. At the RAMP END location, remove the retaining screws for the hold-down bracket. [4]
2. Remove the hold-down bracket by pushing the bracket toward the instrument until it stops, then sliding it right. [5]
3. Repeat Step1 and Step2 to remove the hold-down bracket on the opposite end of the shipping platform. [6]
Reference
i2000/i2000SR Module - Unpack
Prepare to Affix Ramp
Plate
1. At the shipping platform location opposite the RAMP END location, remove the three (3) nuts and washers securing
2.
3.
4.
5.
http://gssweb.add.abbott.com/architect/3a_i2000sr_html/3a_start.htm[5/14/2014 9:48:54 AM]
the shock absorbing feet. [7]
Roll the instrument toward the STOP blocks until it stops. [8]
Lock the instrument wheels in place.
At the RAMP END location, remove the three (3) bolts and washers from the shock absorbing feet. [9]
Remove the feet. [9]
i2000/i2000SR Module - Unpack
Affix Ramp Plate
WARNING
Before proceeding, verify that the instrument wheels are locked and the instrument is in the STOP location.
Failure to do so could result in injury if the instrument moves toward the RAMP END while completing the
next step.
1. Using two (2) bolts, nuts, and washers, affix the ramp plate to the instrument platform.
Remove Instrument
from Shipping Platform
1. Release the wheel locks and roll the instrument toward the RAMP END, and off of the shipping platform.
Position Module
1. Move the processing module to the desired location.
Perform SCC Hardware and Cable
Installation
1. Perform SCC - Hardware and Cable Installation procedure.
i System Service and Support Manual (Version 96661-109) • Copyright 2002, 2006 • ARCHITECT, i2000, i2000SR, ci8200, and c8000 are registered trademarks of Abbott Laboratories. i4000, i6000, i8000, and ARCHITECT ARM are trademarks of Abbott Laboratories. •
Abbott Park, IL 60064 • All rights reserved.
http://gssweb.add.abbott.com/architect/3a_i2000sr_html/3a_start.htm[5/14/2014 9:48:54 AM]
i2000/i2000SR Module - Unpack
http://gssweb.add.abbott.com/architect/3a_i2000sr_html/3a_start.htm[5/14/2014 9:48:54 AM]