REPORT EXTERNAL QUALITY CONTROL COAGUCHEK XS Institution : City : Stichting ECAT Voorschoten Date of control : Monitor : Serial no. : Lot nr. teststrips : Batch no QC set : 02/01/2014 Reference Monitor UP 00112233 510 B221B00 Test results Sample Target %U1 Target Range Measured %D Conclusion 1 2.0 0.6 1.7 - 2.3 1.9 -5.0 OK 2 2.8 0.8 2.4 - 3.2 2.7 -3.6 OK 3 3.6 0.8 3.0 - 4.1 3.4 -5.6 OK 4 4.4 1.3 3.7 - 5.1 4.4 0.0 OK Acceptance range Measured Conclusion > 0.950 0.995 OK Slope : 0.80 - 1.20 1.03 OK Intercept : -0.50 - 0.50 -0.18 OK 4 4 OK 1) Relative uncertainty of the target value. Evaluation Correlation : Correct QC results : Percentage Deviation Correlation Target and Measured value 4.0 Deviation (%) Measured INR 5.0 3.0 2.0 1.0 1.0 2.0 3.0 4.0 5.0 25 20 15 10 5 0 -5 -10 -15 -20 -25 1.0 2.0 Target INR 3.0 4.0 5.0 INR target value FINAL CONCLUSION This CoaguChek XS monitor pass the ECAT criteria for quality control: REMARKS YES Signature + Date: 7 maart 2014 Dr. P. Meijer Director ECAT Foundation ECAT Foundation P.O. Box 107 2254 AG Voorschoten The Netherlands T. +31 71 3030910 E. [email protected] Page 1 of 1