Color compensation Set (Dual Color)

User Manual: Color compensation Set (Dual color) (DIA-DAF(Lc2.0)-005)
Ref.: Dia-DAF(Lc2.0)-005
DIAGENODE
27.08.2014
Color compensation Set
(Dual Color)
Fam
Yellow Dye
For Roche LightCycler® 2.0
www.diagenodediagnostics.com
User Manual: Color compensation Set (Dual color) (DIA-DAF(Lc2.0)-005)
DIAGENODE
PAGE 3
Content
Intended Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Fluorescent Dyes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Product Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Warnings & Precautions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Required Materiel (not supplied). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Create a ccc object. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Performing a Color Hydrolysis probe Color compensation experiment
1. Capillaries preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2. PCR profile and procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Apply a ccc object. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Notice to Purchaser. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Explanation of Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
www.diagenodediagnostics.com
PAGE 4
DIAGENODE
User Manual: Color compensation Set (Dual color) (DIA-DAF(Lc2.0)-005)
Intended Use
With this set, you can create an application-specific Color compensation object using Diagenode
positive control from single color hydrolysis probe reactions.
We recommend to perform Color compensation experiment once a year after the calibration of
the optical part.
Important notes
• It is highly recommended to perform a new compensation file according to the master mix
used.
• The reagents in this Color compensation set have been elaborated especially to be
performed with Diagenode's kits.
Fluorescent Dyes
• FAM (Max Abs 494 nm, Max Em 520 nm)
• Yellow Dye (Max Abs 530 nm, Max Em 549 nm)
Product Description
1. Dye adaptator file primers & double-dye probe: (FAM Dye) 90 µl, green tube.
2. Dye adaptator file primers & double-dye probe: (Yellow Dye) 90 μl, Yellow tube.
3. Positive control 4.103 copies/µl: 200 µl, Blank tube.
4. H2O (PCR grade): 1000 µl, white tube
The product contains primers, probes and a positive control. The master mix is not included.
Storage
Color compensation set must be stored at -20°C until expiry date.
STORE IN DARK PLACE.
Diagenode sa / LIEGE SCIENCE PARK // Rue Bois Saint-Jean, 3 // 4102 Seraing (Ougrée) // Belgium // Phone: (+32) 4 364 20 50 // Mail: [email protected]
User Manual: Color compensation Set (Dual color) (DIA-DAF(Lc2.0)-005)
DIAGENODE
PAGE 5
Warnings and Precautions
If, on receipt, a parcel is found to be damaged or if kits are thawed,
please contact us : [email protected] / +32 4 364 20 52
Material Required (Not Provided)
• Pipettes
• PCR Lc2.0 Roche
• Sterile pipette tips with filters (DNAse/RNAse free)
• Capillaries tubes
• Vortex
• Laminar flow hood (Mix/DNA)
www.diagenodediagnostics.com
PAGE 6
DIAGENODE
User Manual: Color compensation Set (Dual color) (DIA-DAF(Lc2.0)-005)
Create a ccc object
Performing a Color Hydrolysis probe Color compensation experiment
You can create an application-specific Color compensation object using Diagenode positive control
from single color hydrolysis probe reactions.
1. Capillaries Preparation
Prepare 4 replicates Color compensation reactions for each fluorescence dye.
1)Mix Blank: Add the master mix and the Diagenode positive control, but without primers
and probe and adjust volume with water.
2)Mix FAM: Prepare a PCR mix with Diagenode positive control, primers and a FAM-labeled
hydrolysis probe for this control, the master mix, and adjust volume with water.
3)Mix Yellow: Prepare a PCR mix with Diagenode positive control, primers and a yellow dyelabeled hydrolysis probe for this control, the master mix, and adjust volume with water.
We recommend to use with Lc 2.0, the LightCycler® taqman® Master Ref: 04535286001.
PCR 20µl
Blank
FAM
Yellow dye
4 µl
4 µl
4 µl
-
2 µl
2 µl
Diagenode Positive control
5 µl
5 µl
5 µl
Water
11 µl
9 µl
9 µl
Master Mix 5X
Primers & double-dye probe
Important notes :
At the beginning of each run the LightCycler® instrument performs a seek process, which optimize
positionning of the capillaries.
2. PCR profile and procedure
1) Select Run icon
2) In Programs, create your PCR program :
Denaturation Step
Cyles : 1
Target
(°C)
95
Analysis Mode : none
Hold
(hh:mm:ss)
00:10:00
Slope
(°C/s)
Sec Target
(°C)
Step size
(°C)
Step Delay
(cycles)
Acquisition
Mode
20
0
0
0
None
Amplification Step
Cycles : 45
Analysis Mode : None
Target
Hold
Slope
(°C)
(hh:mm:ss)
(°C/s)
Sec Target
(°C)
Step size
(°C)
Step Delay
(cycles)
Acquisition
Mode
95
00 :00 :15
20
0
0
0
None
60
00 :01 :00
20
0
0
0
None
Diagenode sa / LIEGE SCIENCE PARK // Rue Bois Saint-Jean, 3 // 4102 Seraing (Ougrée) // Belgium // Phone: (+32) 4 364 20 50 // Mail: [email protected]
User Manual: Color compensation Set (Dual color) (DIA-DAF(Lc2.0)-005)
DIAGENODE
PAGE 7
Temperature Gradient Step
Cycles : 1
Analysis Mode : Color Compensation
Hold
Target
Slope
(°C/s)
Sec Target
(°C)
Step size
(°C)
Step Delay
(cycles)
Acquisition
Mode
00:00:00
20
0
0
0
None
40
00:00:10
20
0
0
0
None
95
00:00:10
0.2
0
0
0
Continuous
(°C)
(hh :mm
:ss)
95
Cooling Step
Cycles : 1
Analysis Mode : none
Hold
Target
(°C)
(hh :mm
:ss)
40
00:00:30
Slope
(°C/s)
Sec Target
(°C)
Step size
(°C)
Step Delay
(cycles)
Acquisition
Mode
20
0
0
0
None
NB :
Setup :
• Default Channel : 530
• Seek Temperature : 30
• Max. Seek Pos. : 3
• Instrument Type : 6 Ch.
• Capillary Size : 20µl
Sample Editing
1) Select Sample data icon
2) In Analysis Type & Capillary View, complete the table as follows:
Pos
Sample Name
1
Buffer
2
FAM
3
Yellow dye
Repl. Of
Sample Note
NB :
Sample Count : 3
Selected Channels : 530
3) In Analysis Type & Color Comp, complete the table as follows:
Pos
Sample Name
Dominant Channel
1
Buffer
Water
2
FAM
530
3
Yellow dye
560
www.diagenodediagnostics.com
PAGE 8
DIAGENODE
User Manual: Color compensation Set (Dual color) (DIA-DAF(Lc2.0)-005)
Run
1)Place the reagent capillaries on the Carousel in the order listed above, making sure the
blank sample (buffer) is always at position 1.
2)Start the run
Analysis / Back up
When the experiment is finished :
1) Select Color Compensation from the Analysis menu
2)Click Save cc object, then navigate to the folder Special Data\CCC under your user name
and save the file as a ccc object in that folder.
Apply a ccc object
You can now apply the specific ccc object you create to any dual color Hydrolysis Probe experiment
that is performed with FAM / Yellow dye combination, as follows :
1) Click Color compensation in the Run module or in the analysis window
2) Select Color Compensation
3)When a small dialog box opens and dispalys all compensated channels : 530 and 560,
click OK
The experimental data and/or the analysis graphs will now be redrawn using the compensated
data.
Notice to purchaser
LightCycler® 2.0 instrument and LightCycler® taqman® Master are trademarks of Roche.
Diagenode sa / LIEGE SCIENCE PARK // Rue Bois Saint-Jean, 3 // 4102 Seraing (Ougrée) // Belgium // Phone: (+32) 4 364 20 50 // Mail: [email protected]
User Manual: Color compensation Set (Dual color) (DIA-DAF(Lc2.0)-005)
DIAGENODE
PAGE 9
Explanation of symbols
Catalogue number
Batch code
Number of experiments
Temperature limitation
Refer to user manual
Positive control
Manufacturer
yyyy-mm
Expiry Date
Protect from the light
www.diagenodediagnostics.com
PAGE 10
DIAGENODE
User Manual: Color compensation Set (Dual color) (DIA-DAF(Lc2.0)-005)
Diagenode sa / LIEGE SCIENCE PARK // Rue Bois Saint-Jean, 3 // 4102 Seraing (Ougrée) // Belgium // Phone: (+32) 4 364 20 50 // Mail: [email protected]
DIAGENODE
PAGE 11
MA-DAF(Lc2.0)-V1_27_08_14
User Manual: Color compensation Set (Dual color) (DIA-DAF(Lc2.0)-005)
www.diagenodediagnostics.com
Diagenode sa
LIEGE SCIENCE PARK
Rue Bois Saint-Jean, 3
4102 Seraing - BELGIUM
Tel. +32 4 364 20 50
[email protected]
www.diagenodediagnostics.com