BWD881A Powerscope IIA Specifications: VERTICAL AMPLIFIERS CH 1, 2, 3 & 4 - differential amplifiers. CH 5 - single ended amplifier. CH1 x 2 multiplier. Display: 1 to 4 traces from CH 1, 2, 3, 4, 5 or 1x2. Mode: Alternate 0.5msec/div and above. Chopped at 500kHz, 1msec/div and below. Input: 1Mohm±2% and 7.5pF±1pf each side to ground. Calibration: 10 - 40°C 0 - 50°C 0.1V to 200V/div ±3% ±4% 20mV to 50mV/div ±5% ±6% Delay Line: 120nsec delay line is incorporated to display the triggered edge. CH1-4 DIFFERENTIAL AMPLIFIERS Sensitivity: 20mV to 200V/div direct reading in 13 steps. Bandwidth: (8 divisions ref. at 50kHz.) DC or 5Hz to 30MHz -3dB 0.2 to 200V/div. DC or 5Hz to 20MHz -3dB 20 to 100mV/div. Rise Time: (6 divisions references) 12nsec 0.2V to 200V/div. 17nsec 20 - 100mV/div. Max Input: ±1,000V or 660Vrms, 3kV peak overload for 1 sec. CH4 INPUT OFFSET Multi-turn control providing the following offset ranges: 20mV to 10V/div. >±200V 20V to 200V/div. >±3,600V Common Mode Rejection: 86dB to 1kHz 56dB at 1MHz Max. Common Mode Input 350Vrms or ±500V, 20mV - 10V/div. 660Vram or ±1,000V 20V - 200V/div. CH1 x 2 MULTIPLIER Sensitivity: CH1 x 2 x 1/5, ie. 4 div on CH1 and 6 div on CH2 will produce a signal 24/5 = 4.8 divisions. Bandwidth: DC or 10Hz to 12MHz, -3dB. Accuracy: Better than 10% up to 5MHz with AC signal applied to one channel and DC to the other. CH5/EXT. TRIGGER SINGLE SIDED AMPLIFIER (selected as alternative to CH4) Sensitivity: 0.2V to 10V/div in 6 steps. Accuracy: ±3%, 10 to 40°C ±4%, 0 to 50°C. Bandwidth: DC or 5Hz to 50MHz, -3dB. Rise Time: 7nsec. Input: 1Mohm±2%, 7pF±2pF. Max Input: ±1,000V (overload condition). X-Y OPERATION Single channel identical sensitivity X-Y display. Bandwidth and calibration of Y channel identical to vertical display. X channel - DC to 1MHz, -3dB. Calibration: ±5% Phase Shift: less than 1° - DC to 200kHz. TIME BASE GENERAL DETAILS Ranges: 0.2usec to 1 sec/div in 21 steps. A vernier covers the range between steps and extends the range to approx. 5 sec/div. Magnification: x1, x10. Max sweep speed 20nsec/div. Calibration: (measured over centre 8 divs). 10 to 40°C x1 mag ±3% x10 mag ±5% 0 to 50°C x1 mag ±4% x10 mag ±6% 1 sec/div range only - x1mag ±6%, x10mag ±8% over 0 to 50°C CRT: 8 x 10 div (1 div = 1cm), 10kV EHT. P31 aluminised phosphor, internal edge illuminated graticule. Blue filter. Probe Cal: Approx 1kHz rectangular wave, positive going with respect to ground 10V pkpk, 1% accuracy. Beam Find: Push button returns trace within CRT limits and overrides intensity control. Z Modulation: +5V with respect to ground will blank trace at normal intensity. Bandwidth DC to greater than 10MHz. Input 10kohm. Max input ±30V. Front Panel Input Sockets: Special snap fit shrouded high voltage. BWD A8 adapter available to convert input to standard BNC. Rear Panel Output Signals: Each individual channel at 50mV/div ±5% is available from BNC sockets with approx 500 ohms source impedance. Combined signals as displayed on the CRT at approx 50mV/div is available from BNC socket with approx 150 ohms source impedance. TRIGGERING Source: CH1, 2, 3, 4, 5, Line Frequency or Phase Marker Pulse. Coupling: DC or AC (5Hz, -3dB nom.) LF Filter: Switch selection of a DC to 400Hz low pass filter for stable triggering from very noisy low frequency waveforms. Polarity: + or - with level select over 8 div of CRT deflection. Modes:Auto, normal or single sweep with manual or triggered initiation. Ready lamp indicates state of trigger under all conditions. Sensitivity: CH1 - 4 - 0.5 div or greater, DC to 2MHz increasing to 3 div at 30MHz. CH5 - 0.5div or greater DC to 10MHz increasing to 3 div at 50MHz. ENVIRONMENTAL Temperature: Operating 0 to 50°C Storage -55°C to +75°C. Humidity: 95% RH non-condensing 0 to 40°C, 50% RH non-condensing from 0 to 50°C. CH5 - AS EXT. TRIGGER Sensitivity range extends from 100mV to 100V pk-pk with 1:1 probe and 1V to 1,000V pk-pk with a 10:1 probe. Trigger sensitivity as CH5 amplifier. PHASE MARKER TRIGGER The phase marker pulse can be selected to trigger the time base. This enables the time base speed to be expanded until the 1° wide phase marker is 1 div wide to obtain high resolution measurements of phase and pulse jitter. PHASE GENERATOR Range: Selectable in 1° steps from 0 to 359° by individual UP/DOWN buttons. Single degree steps occur when a button is momentarily depressed. Continuous counting occurs at approx 20°/sec when either button is held depressed. Readout: 10mm LED display in degrees. Frequency Range: 15Hz to 2kHz. Automatic operation, response time less than 0.5 secretary for a 10% frequency change. Phase Reference Source: CH1, 2, 3, 4 and 5 or zero cross over of input AC power (within 1° at 50-60Hz). 0° Reference Selector: 0° reference and the phase marker pulse switchable in 60° steps from 0° to 300°. 0° Reference Control: Moves the zero reference to any position on a displayed line frequency waveform. 100usec to over 20msec delay. Marker Pulse Jitter: Less than 1° from a stable noise free source. Phase Resolution: Less than 0.5° between similar pulses in multi-phase circuits. Noise Filter: 400Hz low pass for reference signals with multiple zero crossing. POWER AC: 90 to 264Vrms, 45 to 440Hz, 40W max. DC: 100 to 250V IEC 6A power socket on rear panel. DC: 20 to 32V (1.5A at 24V) using the optional PS1 power supply adapter. Battery: Optional BP6 battery pack - also requires the optional PS1. SAFETY Instruments closely conforms to IEC348, UL and VDE. The entire operating section controls, panel etc are insulated and the front panel sockets are deeply recessed to prevent contact with grounded metal surfaces. ACCESSORIES INCLUDED 2 off 2 off 1 off 1 off BWD P90 1:1 probes BWD P91 10:1 probes BWD P95 10:1/1:1 probe Instruction and Application Manual WEIGHT Nett: Shipping: PS1: BP6: 9.4kg 12kg, for options add: 0.5kg 7.5kg DIMENSIONS Instrument Face: 355 (w) x 180 (h) Overall: 395 (w) x 270 (h) x 470 (d) Shipping Carton: 440 (w) x 340 (h) x 530 (d) Specifications subject to change without notice 58 Geddes Street, PO Box 298, Mulgrave Victoria, 3170, AUSTRALIA Tel: (+61-3) 9582-7333, Fax: (+61-3) 9560-1164 E-mail: [email protected], Internet: www.mcvan.com BWD881A Powerscope IIA Probes & Accessories: P90R and P90B 1:1 Probes Non-attenuating probe primarily intended for use in pairs for differential measurement of low level signals superimposed on high voltages whilst retaining the common mode rejection of the Powerscope amplifiers. Bandwidth: Up to 25MHz. Rise Time: From 14nsec. Input Resistance: 1Mohm of Powerscope input. Input Capacitance: Approx 100pF with twisted input leads. Input Voltage: Max 1,000Vdc/660Vrms, 3kV peak for 1 second max. Input Protection: Inbuilt fuse 1,000V/100mA Cable Length: 2 metres (approx 6.5ft) Cable Colour: Black (P90B) or Red (P90R) Probe Connection: Detachable grabber opening to 10mm and shrouded 4mm plug. P91 10:1 Probes A 10:1 attenuating probe for use with large amplitude signals. Probes may be used in pairs for high level differential measurement. Each probe is adjustable to match CMR response. Bandwidth: 40MHz., -3dB Rise Time: Less than 9nsec. Input Resistance: 5Mohm (with Powerscope) Input Capacitance: Approx 13pF. Compensation: Adjustable for resistance and capacitance. Input Voltage: Max 2,000Vdc/1.5kVrms, 3kV peak for 1 second max. Cable Length: 2 metres (approx 6.5ft) Probe Connection: Detachable grabber opening to 10mm and shrouded 4mm plug. P95 and P86 Miniature Probes. The P95 is a switchable 1:1 and 10:1 probe and the P86 a fixed 100:1 attenuating probe. These passive probes are of modular construction which facilitates simple field replacement of the consumable parts. Both are supplied with 1.5metres (approx 5ft) of cable. P95 (x1) P95 (x10) P86 (x100) Bandwidth: 8MHz 100MHz 100MHz Input Resistance: 1Mohm 10Mohm 100Mohm Input Capacitance: 47pF 15pF 6.5pF Max Input Voltage: 360V 360V 1,200V Rise Time: 45nsec 3.5nsec 1.4nsec Compensation Range: 10-60pF 10-60pF 5-50pF Current Probes Truly portable clamp-round probes that you measure dc, ac and ac on dc currents. Ideal for power measurements when using the CH1x2 multiplier with the current probe connected to CH1 and the voltage waveform to CH2. P30 P82 Measurement Range: ±20A 20/200A (2 ranges) Accuracy: ±1%±2mA 2% of output Output: 100mV/A ±2V = 20A/200A Frequency Response: DC to 100kHz DC to >1kHz Output Impedance: 100 ohms nominal 125 ohms nominal. P87 High Voltage Probe A 1,000:1 attenuating probe for use with large amplitude signals. Features include very high input impedance, wide bandwidth and HF compensation. Bandwidth: DC to 5MHz Input Resistance: 500Mohm Input Capacitance: 3pF Working Voltage: 12.5kV or peak AC. Other Accessories include: A8 BNC to Powerscope socket adapter. PS1 Power Supply Adapter for operation from 20 - 32Vdc. BP6 Battery Pack (requires PS1). PS1 - 20-32Vdc Power Supply Converter BP6- 24V Battery Pack with inbuilt charger 58 Geddes Street, PO Box 298, Mulgrave Victoria, 3170, AUSTRALIA Tel: (+61-3) 9582-7333, Fax: (+61-3) 9560-1164 E-mail: [email protected], Internet: www.mcvan.com Specifications subject to change without notice File: BWD881A Data Sheet Nov 01
© Copyright 2024 ExpyDoc