Clin Chem Lab Med 2014; aop Letter to the Editor Sutirtha Chakraborty*, Sidhartha Sankar Ghosh, Anupam Das, Prasad Sawant and Anders Kallner Can EDTA, EDTA-fluoride, and buffered citrate tubes be used for measurement of HbA1c on the Bio-Rad D10? Keywords: analysis of variance components; diabetes monitoring; different tubes; emergency co-requests; error grid. DOI 10.1515/cclm-2014-0644 Received June 19, 2014; accepted July 6, 2014 To the Editor, The measurement of HbA1c in blood has become the gold standard for long-term control of the glycemic state of diabetic patients. The concentration of blood glucose indicates the real-time glycemic status at the time of sampling, whereas the HbA1c concentration indicates the glycemic status over the preceding 8–12 weeks. In clinical practice, both glucose and HbA1c concentrations are important in diabetes monitoring because they provide complementary information. Recently, HbA1c has also been recommended in the diagnosis of diabetes [1]. Simultaneous requests of HbA1c and glucose are common, and alone or together may trigger a revision of the therapeutic regime. Post-prandial glucose concentrations significantly contribute to HbA1c concentration. Therefore, post-prandial glucose requests are required if the therapeutic targets of HbA1c are not met [2]. Because *Corresponding author: Dr. Sutirtha Chakraborty, Department of Biochemistry, Peerless Hospital and B. K. Roy Research Centre, Kolkata 700094, India, Phone: +91 33 24622462 (extension 2204), E-mail: [email protected]; [email protected] Sidhartha Sankar Ghosh: Department of Biochemistry, Peerless Hospital and B. K. Roy Research Centre, Kolkata, India Anupam Das: Department of Quality Assurance, Peerless Hospital and B. K. Roy Research Centre, Kolkata, India Prasad Sawant: Asian Institute of Oncology Pvt Ltd, Department of Laboratory Medicine, Mumbai, India Anders Kallner: Department of Clinical Chemistry, Karolinska University Hospital, Stockholm, Sweden HbA1c does not require a fasting patient, co-request becomes a convenient and efficient procedure in rational diabetes monitoring. Glucose concentration is measured in blood, serum, or plasma, whereas for HbA1c concentration, EDTA blood is recommended [3]. To minimize glycolysis, in vitro fluoride is added to samples for glucose concentration measurements. This addition gives full effect after about 2 h, whereas buffering the EDTA-fluoride to pH of approximately 5.5 [4, 5] has been shown to instantaneously inhibit glycolysis. We have only found two studies that validate the use of anticoagulants other than EDTA in the measurements of HbA1c [6, 7]. These were limited to 6 and 30 patient samples, respectively. Although claiming equivalence of results using EDTA and EDTA-fluoride tubes, no statistics were presented. We investigated EDTA anticoagulated blood with added fluoride and buffered fluoride for measuring HbA1c concentration. This is a methodological change and thus requires [8] validating if the “in-house measurement procedure” is fit for purpose, i.e., to describe the degree of agreement among results obtained from EDTA, EDTAfluoride, and buffered EDTA-fluoride tubes. The study cohort included in-patients and out-patients who were either diabetic or screened for glycemic impairment. This ensures that the cohort includes samples from the entire measuring interval although with an expected predominance of increased values. Samples were collected in vacuum tubes; Vacutainers® were Potassium EDTA and Sodium Fluoride-Sodium EDTA (Becton Dickinson, Franklin Lakes, NJ, USA) and Terumo Venosafe® – Glycemia Buffered Fluoride Sodium EDTA tubes (Terumo, Belgium). The latter were buffered by citrate and thus contain two potent calcium complex binders. HbA1c concentrations were measured using a Bio-Rad D10 cation exchange high-performance liquid chromatography (HPLC) platform (Bio-Rad Laboratories, Hercules, Brought to you by | provisional account Unauthenticated | 10.248.254.158 Download Date | 8/8/14 5:13 AM 2 Chakraborty et al.: Can glucose tubes be used for measurement of HbA1c? 140 120 100 HbA1c mmol/mol CA, USA) using the Bio-Rad calibrators with assigned values traceable to the IFCC Reference Measurement Procedure. The results were reported in relative substance concentration units, mmol/mol. Venous blood, 2 mL, was collected from 104 patients. Owing to limited amounts of material, the study was made in two sets. In the first 104 patients, results were compared after collection in EDTA and EDTA-fluoride tubes (set I). In the second, another set of results from 104 patient were obtained in EDTA blood and were compared with those from buffered EDTA-fluoride tubes (set II). All pairs of samples were processed on the same day. The validation was focused on estimating any bias between results obtained in the different tubes and the imprecision of measurement procedures. The bias was estimated using regression analysis [9] and absolute and relative difference graphs. Excel spreadsheet programs were used for regression and difference analyses. Nonparametric tests were performed, when appropriate, using Sigma Plot (v. 12.5 Systat, Germany). A detailed imprecision evaluation was completed using “analysis of variance components” (ANOVA) as described in CLSI EP15 [10] and Kallner [11]. The clinical importance of the differences was evaluated using error grids comprising A-, B-, and C-zones [12]. Significance was considered at p < 0.05. The distribution of the HbA1c concentrations in the cohorts was slightly skewed to higher values as expected and failed the Shapiro-Wilks normality test. The medians of the results of the two sets, collected in EDTA, were not statistically different using Mann-Whitney rank sum test. The differences between the paired results also did not pass the normality test, and thus, the significance of pairwise differences was tested using Wilcoxon’s signed rank test. The results and comparisons are shown as box-plots in Figure 1, which also shows the skewness of the data. There were no suspected outliers visible in the scatter plots, and the coefficients of determination (r2) were high ( > 0.99). The average difference between the observations in set I was 0.5 mmol/mol, and in set II, –0.5 mmol/mol, whereas the median difference between the measurements was 0.0 mmol/mol in both sets. Although the numerical differences were small, a statistical significant difference was shown by Wilcoxon’s signed rank test between the EDTA tubes and those with additives. Regression functions were calculated using Deming regression with a relative variance of independent and dependent variables (λ) set to 1. The slope was 1.00 and 0.99, and the intercept, 0.14 and –0.04 in set I and set II, respectively. We described the clinical performance in terms of error grids [12]. In this approach, the widths of various 80 60 40 20 0 Set 1 A EDT o +Flu Set 1 Set 2 A EDT o f Flu +buf Set 2 Sampling tube Figure 1 Box-plots of the results in the investigated tubes. Boxes represent the interquartile interval, the whiskers the 5th and 95th percentiles. Outside the whiskers, each observation is shown. The solid line in the box represents the median, and the dotted, the average. The difference between these central measures illustrates the skewness of the data distribution. zones of accepted results are defined and the number of observations in each zone is counted. The A-zone is defined symmetrically around the equal line as a ± 5% deviation. It should ideally contain 95% of the observation. For set I, 92% of the observations were within the A-zone, and for set II, 97%. The remaining observations were in the B-zone, defined as ± 10%, symmetrical around the equal line and thus including the A-zone. There were none outside its limits, i.e., in the C-zone. The imprecision study comprised measuring five samples each day for 5 days using a large enough sample, appropriately aliquoted from one and the same patient for each set. The evaluation was done using ANOVA components. This has the advantage of offering within- and pure between-series variances [10, 11]. Their sum (combined uncertainty) corresponds to within-laboratory variance. The detailed results of the analysis of the variance components are shown in Table 1. The imprecision was of the same order of magnitude independent of the tube and within the target total imprecision of 3%. The minimal difference [13] between two successive measurements at Brought to you by | provisional account Unauthenticated | 10.248.254.158 Download Date | 8/8/14 5:13 AM Chakraborty et al.: Can glucose tubes be used for measurement of HbA1c? 3 Table 1 Imprecision of measurements. Sample set I EDTA Within Pure between Laboratory Concentration mmol/mol % 0.47 0.19 0.51 0.9 0.4 1.0 53.0 Sample set II EDTA-fluoride mmol/mol % 0.45 0.23 0.50 0.8 0.4 0.9 54.0 a level of probability of 95% is thus about 1.2 mmol/mol (2.2% at the tested concentration). The data analysis shows that there is a statistically significant difference between the results attributable to the tubes. This difference, estimated as a mean difference by regression analysis or at specified concentration, is of the same order of magnitude, or less, than the minimal significant detectable difference [13] and will therefore not have any clinical implications. Sacks et al. [14] discussed the analytical goals for HbA1c measurements. The presentation of data is not fully comparable, but the recommendation suggests an intra-laboratory coefficient of variations of < 2%; our corresponding value was < 1.5% for both types of tubes in our study (Table 1). Expressing the relative performance of the methods as the deviation from the equal line in terms of different zones makes the performance understandable outside the analytical laboratory and metrology, e.g., among clinicians. The high correlation coefficient and proportion within the A-zone and lack of observations in the C-zone are measures of equality of individual results. The statistical power of the EP15 may not be sufficient to create an imprecision profile, but the degrees of freedom allow a comparison of the combined uncertainty of the methods. The ANOVA components allows the isolation of within (repeatability) and between (reproducibility) variation. The uncertainty contribution in our study was mainly from the repeatability component, i.e., the reproducibility was high. It is not unusual to observe that the between-series variance is smaller than that within series in measurement procedures where the measuring system is stable, i.e., calibration is reproducible and the reagents stable. Using ANOVA components is therefore desirable in validating and verifying measurement procedures. The present study therefore recommends that improvements of the performance should focus on the repeatability of the measurements. Simultaneous requests for HbA1c and glucose concentration measurements are common. It is demonstrated that HbA1c can be measured equally well using hemolysates from EDTA mmol/mol % Buffered EDTA-fluoride mmol/mol % 0.58 0.45 0.74 1.0 0.8 1.3 57.5 0.58 0.16 0.60 1.0 0.3 1.1 56.9 EDTA tubes, EDTA-fluoride, and buffered EDTA-fluoride tubes on the Bio-Rad D10. This will contribute to rational treatment of patients, savings of materials, and reduction of inconveniences for patients and staff. Subsequent measurement of HbA1c concentration helps distinguish between reactive and diabetic hyperglycemia and can safely be requested from the original glucose tube. The present study is limited to one measurement procedure, and the compatibility of different glucose tubes for HbA1c measurements using non-HPLC methods needs further verification. Author contributions: All the authors have accepted responsibility for the entire content of this submitted manuscript and approved submission. Financial support: None declared. Employment or leadership: None declared. Honorarium: None declared. Competing interests: The funding organization(s) played no role in the study design; in the collection, analysis, and interpretation of data; in the writing of the report; or in the decision to submit the report for publication. References 1. The Diabetes Control and Complications Trial Research Group. The effect of intensive treatment of diabetes on the development and progression of long term complications in insulin-dependent diabetes mellitus. N Engl J Med 1993;329:977–86. 2. Monnier L, Lapinski H, Colette C. Contributions of fasting and postprandial plasma glucose increments to the overall diurnal hyperglycemia of type 2 diabetic patients. Diabetes Care 2003;26:881–5. 3. Jeppsson J-O, Kobold U, Barr JR, Finke A, Hoelzel W, Miedema K, et al. Approved IFCC reference method for the measurement of HbA1c in human blood. International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) Network of Reference Laboratories for HbA1c. Clin Chem Lab Med 2002;40:78–89. 4. Gambino R, Piscitelli J, Ackattupathil TA, Theriault JL, Andrin RD, Sanfilippo ML, et al. Acidification of blood is superior to sodium fluoride alone as an inhibitor of glycolysis. Clin Chem 2009;55:1019–21. Brought to you by | provisional account Unauthenticated | 10.248.254.158 Download Date | 8/8/14 5:13 AM 4 Chakraborty et al.: Can glucose tubes be used for measurement of HbA1c? 5. Fobker M. Stability of glucose in plasma with different anticoagulants. Clin Chem Lab Med 2014;52:1057–60. 6. Mailankot M, Thomas T, Praveena P, Jacob J, Benjamin JR, Vasudevan DM. Various anticoagulants and fluoride do not affect HbA1C level. Indian J Clin Biochem 2012;27:209. 7. Boonlert W, Narakantha P, Onseng S, Yamphochai S, Nuanmuang N, Mohprasit K. Hemoglobin A1c levels obtained from EDTA, lithium heparin, lithium heparin plus glyceraldehydes, and sodium fluoride blood samples. Songklanagarind Medical Journal 2010;28:107–16. Abstract cited in Available from: http://thailand.digitaljournals. org/index.php/SOMJ/article/view/4241 and the original cited in http://thailand.digitaljournals.org/index.php/SOMJ/article/ viewFile/4241/3791. Accessed on 2 June, 2014. 8. EN/ISO 15189. Medical laboratories – requirements for quality and competence. Geneva: ISO, 2011. 9. EP9. NCCLS. Method comparison and bias estimation using patient samples: approved guideline, 2nd ed. NCCLS document EP9-A2 (ISBN 1-56238-472-4). Wayne, PA: NCCLS, 2002. 10. EP15. CLSI. User verification of performance for precision and trueness: approved guideline, 2nd ed. CLSI document EP15-A2 (ISBN 1-56238-574-7). Wayne, PA: CLSI, 2005. 11. Kallner A. Laboratory statistics. Handbook of formulas and terms. New York: Elsevier, 2014:54–6. 12. EP27. CLSI. How to construct and interpret an error grid for quantitative diagnostic assays: approved guideline. CLSI document EP27-A (ISBN 1-56238-853-3). Wayne, PA: CLSI, 2012. 13. C51-EP29. CLSI. Expression of measurement uncertainty in laboratory medicine: approved guideline. CLSI document C51-A (ISBN 1-56238-787-1 [print]; ISBN 1-56238-788-X [electronic]). Wayne, PA: CLSI, 2012. 14. Sacks DB, Arnold M, Bakris GL, Bruns DE, Horvath AR, Kirkman MS, et al. Executive summary: guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Clin Chem 2011;57:793–8. Brought to you by | provisional account Unauthenticated | 10.248.254.158 Download Date | 8/8/14 5:13 AM
© Copyright 2024 ExpyDoc