RECD-Measurements with Connexx 6.3 and Unity 2

RECD-Measurements
with Connexx 6.3 and Unity 2
Vera Spitzlei, Corporate Audiology
Siemens Audiologische Technik GmbH
For internal use only / Copyright © Siemens AG 2009. All rights reserved.
Overview
ƒ Why measure RECD?
ƒ RECD with Unity 2
ƒ RECD with Unity via Connexx ClinicalFit
ƒ RECD in Unity: Extended Options
ƒ RECD: Possible Pitfalls
Page 2
Nov-09
Vera Spitzlei
H WS AU MB A
Overview
ƒ Why measure RECD?
ƒ RECD with Unity 2
ƒ RECD with Unity via Connexx ClinicalFit
ƒ RECD in Unity: Extended Options
ƒ RECD: Possible Pitfalls
Page 3
Nov-09
Vera Spitzlei
H WS AU MB A
Why measure RECD?
Anatomical terms
ƒ Every ear canal has an
individual shape and size!
Î Coupler or ear simulator can
only provide an estimate /
approximation of the residual
ear canal volume and resulting
acoustics at the eardrum
ƒ Esp. in children, auricle and ear
canal are smaller, more straight
and more narrow than in adults
Î Resonance peaks are shifted
towards higher frequencies
Page 4
Nov-09
Vera Spitzlei
H WS AU MB A
Why measure RECD?
resonance peaks in ear canal
shifted to higher frequencies!
RECD according to age
REUG according to age
REUG 9t14m
REUG 14t24m
RECDd0t12(CXX)
RECDd49t60(CXX)
REUG 24andup
30
30
25
25
20
20
15
L[dB]
L[dB]
REUG 0t4m
10
5
0
100
-5
RECDd25t48(CXX)
15
10
5
1000
10000
0
-5100
f[Hz]
Page 5
RECDd13t24(CXX)
RECDd60andup(CXX)
1000
10000
f[Hz]
Nov-09
Vera Spitzlei
H WS AU MB A
Why measure RECD?
This might contribute to
REIG gap@FOG50dB,
related to adult values (>60 months)
a higher sound level
3 months
12 months
24 months
36 months
48 months
25
at the eardrum of
20
more than 20 dB!
L[dB]
15
10
5
0
100
Danger of
over-amplification!
Page 6
Nov-09
1000
10000
-5
-10
f[Hz]
Vera Spitzlei
H WS AU MB A
Why measure RECD?
Measuring options for verification under real-ear (In-Situ) conditions
recommended stimuli for NAL-NL1 and DSL:
ISTS (speech signal), ILTASS (un-modulated), Icra1 (un-modulated), Icra7 (modulated)
RECD measurement
In-Situ measurement
REIG measurement
coupler measurement
Page 7
Nov-09
OEG measurement
Vera Spitzlei
H WS AU MB A
Why measure RECD?
Measuring options for verification under real-ear (In-Situ) conditions
recommended stimuli for NAL-NL1 and DSL:
ISTS (speech signal), ILTASS (un-modulated), Icra1 (un-modulated), Icra7 (modulated)
RECD measurement
In-Situ measurement
REIG measurement
coupler measurement
Page 8
Nov-09
OEG measurement
Vera Spitzlei
H WS AU MB A
Why measure RECD?
RECD
(Real-Ear-to-Coupler Difference / Moodie, Seewald et. al. 1994)
RECD describes the difference of an acoustic response, measured in a
2cc-coupler and in an individual’s ear canal, using the same sound
source, as a function of frequency in dB SPL
Î The individual residual ear canal volume can be taken into account
without the need of a complete In-Situ measurement for each fitting
Î Calculated parameters can be stored
Î Least possible strain for the patient
Page 9
Nov-09
Vera Spitzlei
H WS AU MB A
Overview
ƒ Why measure RECD?
ƒ RECD with Unity 2
ƒ RECD with Unity via Connexx ClinicalFit
ƒ RECD in Unity: Extended Options
ƒ RECD: Possible Pitfalls
Page 10
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2
UNITY™ 2
Page 11
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2
Start from
NOAH module
Start from
Unity
environment
Page 12
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2
Client
list
Change ear
sides
Connexx
module
Session
Save
Print
list
fitting
fitting
session session
Page 13
Open Ear
Unaided
Spectrum
measurement
(Open ear Insertion
gain)
Gain
RECD
Probe tube
calibration
Sound
Aided
Input/
Mapping
Measurement Output
(InSitu and diagram
coupler)
End program
Occluded
Occlusion
measurement
effects
SPL / Gain
view
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2
RECD
Probe tube
calibration
End program
Page 14
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2
Probe tube
calibration
Page 15
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2:
Probe Tube Calibration
A calibration of the probe tube
must be carried out prior to a realear measurement.
It should be repeated every time
the probe tube is replaced.
Page 16
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2
RECD
Page 17
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2
Page 18
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2
Unity 2 supports two different methods for obtaining RECD:
ƒ With hearing instruments in free soundfield
ƒ Through insert earphones
(e.g. with EARtones ER3A / ER5A or using the Unity In-Situ – Headset)
Page 19
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 – Freefield (with HI)
RECD measurement via Freefield (with HI):
1. In-Situ measurement
ƒ
Gain measurement via HI
at the eardrum
2. Coupler measurement
ƒ
Gain measurement via HI
in 2cc-coupler
Î Resulting difference = Real-Ear-to-coupler Difference (RECD) !
Page 20
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 – Insert Earphones
RECD measurement via Insert Earphones
Options:
Insert Earphones for RECD Æ
Insitu headset
or Insert Earphones
(EARtone ER3A / ER5A)
Unity Æ Settings
Æ Devices Æ Settings
Page 21
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 – Insert Earphones
Stage 1: Coupler Measurement
Page 22
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 – Insert Earphone A
1a) Application of test signal to 2cc HdO - coupler (HA2)
via Insert Earphone
(EARtone ER3A/ER5A)
Page 23
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 – Insert Earphone A
1a) Application of test signal to 2cc HdO - coupler (HA2)
via Insert Earphone
(EARtone ER3A/ER5A)
Page 24
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 – Insert Earphone B
1b) Application of test signal to 2cc HdO - coupler (HA2)
via In-Situ headset (Unity)
Page 25
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 – Insert Earphone B
1b) Application of test signal to 2cc HdO - coupler (HA2)
via In-Situ headset
(Unity)
Page 26
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 – Insert Earphone A
2a) Determine the appropriate tube length for the probe
microphone
Page 27
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 – Insert Earphone B
2b) Attach an ear tip to the end
of the handle. The handle
contains both, the receiver
tube and the probe tube.
The probe outlet of the handle
tube should always be
elongated with the short, 5mm
tube to get the opening closer
to the eardrum.
Page 28
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 – Insert Earphone A
3a) Placement of probe tube in the ear canal
Æ measuring sound pressure level
directly at the eardrum
Page 29
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 – Insert Earphone A
4a) Insertion of earmold
(+ connection of Insert Earphone)
Æ Measurement setup more complex
and time-consuming
BUT
Æ Measured residual volume
DOES consider the individual
earmold
Page 30
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 – Insert Earphone B
4b) Insertion of Insitu-Headphone
Æ Measurement faster, no separate
In-Situ – tube necessary
BUT
Æ measured residual volume
does NOT consider the individual
earmold
Page 31
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 – Insert Earphone
Stage 2:
In-Situ measurement
5) Application of reference signal
6) Individual RECD measurement
Page 32
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 – Insert Earphone
Stage 2:
In-Situ measurement
Æ resulting curve
= Real-ear-to-CouplerDifference
= RECD !
Page 33
Nov-09
Vera Spitzlei
H WS AU MB A
Overview
ƒ Why measure RECD?
ƒ RECD with Unity 2
ƒ RECD with Unity via Connexx ClinicalFit
ƒ RECD in Unity: Extended Options
ƒ RECD: Possible Pitfalls
Page 34
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity via Connexx ClinicalFit
There are two ways to insert measured RECD values in ClinicalFit:
1. By clicking in the RECD grid once, a dialog comes up which enables
the insertion of individual RECD for each frequency available by
putting in numeric values.
2. By double clicking in the RECD grid to confirm the individual RECD for
each frequency available. In this case, a tool tip gives additional
information on which value (in dB) has been selected. Additionally,
the inserted values are automatically displayed as numeric values in
the dialog for manual insertion.
Page 35
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit
Page 36
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit
Parameters:
ƒ HI Adaptive System: Initial / Off
ƒ HI Microphone System: Directional / Omnidirectional
ƒ LO Max (Maximum Level Output, default: 105 dB SPL)
ƒ Test Signal (default: White Noise)
ƒ LI (Level Input, default: 65 dB)
Response Curve display list (show / print / delete)
Tube
Calibration
Start/Stop
Page 37
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit:
Probe Tube Calibration
Page 38
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit:
Probe Tube Calibration
A calibration of the probe tube
must be carried out prior to a realear measurement.
It should be repeated every time
the probe tube is replaced.
Page 39
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit –
Freefield (with HI)
Page 40
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit –
Freefield (with HI)
RECD measurement via Freefield (with HI):
1. In-Situ measurement
ƒ
Gain measurement via HI
at the eardrum
2. Coupler measurement
ƒ
Gain measurement via HI
in 2cc-coupler
Î Resulting difference = Real-Ear-to-coupler Difference (RECD) !
Page 41
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit –
Insert Earphones
Page 42
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit –
Insert Earphones
RECD measurement via Insert Earphones
Options:
Insert Earphones for RECD Æ
Insitu Headphone
or Insert Earphone
(EARtone ER3A / ER5A)
Unity Æ Settings
Æ Devices Æ Settings
Page 43
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit –
Insert Earphones
Stage 1: Coupler Measurement
Page 44
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit –
Insert Earphone A
1a) Application of test signal to 2cc HdO - coupler (HA2)
via Insert Earphone
(EARtone ER3A/ER5A)
Page 45
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit –
Insert Earphone A
1a) Application of test signal to 2cc HdO - coupler (HA2)
via Insert Earphone
Page 46
Nov-09
(EARtone ER3A/ER5A)
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit –
Insert Earphone B
1b) Application of test signal to 2cc HdO - coupler (HA2)
via In-Situ Insert Earphone (Unity)
Page 47
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit –
Insert Earphone B
1b) Application of test signal to 2cc HdO - coupler (HA2)
via In-Situ Insert Earphone (Unity)
Page 48
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit –
Insert Earphone A
2a) Determine the appropriate tube length for the probe
microphone
Page 49
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit –
Insert Earphone B
2b) Attach an ear tip to the end
of the handle. The handle
contains both, the receiver
tube and the probe tube.
The probe outlet of the handle
tube should always be
elongated with the short, 5mm
tube to get the opening closer
to the eardrum.
Page 50
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit –
Insert Earphone A
3a) Placement of probe tube in the ear canal
Æ measuring sound pressure level
directly at the eardrum
Page 51
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit –
Insert Earphone A
4a) Insertion of earmold
(+ connection of Insert Earphone)
Æ Measurement setup more complex
and time-consuming
BUT
Æ Measured residual volume
DOES consider the individual
earmold
Page 52
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit –
Insert Earphone B
4b) Insertion of Insitu-Headphone
Æ Measurement faster, no separate
In-Situ – tube necessary
BUT
Æ measured residual volume
does NOT consider the individual
earmold
Page 53
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit –
Insert Earphones
Stage 2:
In-Situ – Measurement
5) Application of reference signal
6) Individual RECD measurement
Page 54
Nov-09
Vera Spitzlei
H WS AU MB A
RECD with Unity 2 via Connexx ClinicalFit
Stage 2:
In-Situ – Measurement
red curve: measured RECD
blue curve: statistal RECD
Page 55
Nov-09
Vera Spitzlei
H WS AU MB A
Overview
ƒ Why measure RECD?
ƒ RECD with Unity 2
ƒ RECD with Unity via Connexx ClinicalFit
ƒ RECD in Unity: Extended Options
ƒ RECD: Possible Pitfalls
Page 56
Nov-09
Vera Spitzlei
H WS AU MB A
RECD in Unity: Extended Options
Manual access to DSL v5 dialog
Page 57
Nov-09
Vera Spitzlei
H WS AU MB A
RECD in Unity: Extended Options
DSL v5 Configuration:
Insert
ƒ (developmental) Age
ƒ Transducer used for Assessment
ƒ RECD type
ƒ Number of HI Channels
Page 58
Nov-09
Vera Spitzlei
H WS AU MB A
RECD in Unity: Extended Options
Æ DSL v5 Configuration
Æ Advanced Settings
Page 59
Nov-09
Vera Spitzlei
H WS AU MB A
Verification in Unity:
Insert measured RECD in Unity
Æ DSL v5 Configuration
Æ Advanced Settings
Æ RECD options:
Predicted / Measured / Manual
Æ RECD “Manual”:
Insert values for Left / Right ear
Page 60
Nov-09
Vera Spitzlei
H WS AU MB A
Overview
ƒ Why measure RECD?
ƒ RECD with Unity 2
ƒ RECD with Unity via Connexx ClinicalFit
ƒ RECD in Unity: Extended Options
ƒ RECD: Possible Pitfalls
Page 61
Nov-09
Vera Spitzlei
H WS AU MB A
RECD: Possible Pitfalls
Insecure Measurements /
Possible Pitfalls :
1. Capping at 2 - 3 kHz
2. Negative values at lower
frequencies (-1 to -9 dB)
3. Negative values at lower
frequencies (-10 to -15 dB)
1. Insertion of Probe Tube not
deep enough
2. Acoustical leak / venting
4. Positive values at low and
medium frequencies
Page 62
Nov-09
3. Grommets or perforation
4. Eardrum problems / effusion
Vera Spitzlei
H WS AU MB A
Summary
RECD with Unity 2 and Connexx 6.3 ClinicalFit:
ƒ Use of individually measured or
statistical (age-related) RECD
Î Optimal support especially for
pediatric fittings
ƒ Perfect amendment through
verification options via Unity
Î Safe and effective fitting!
Page 63
Nov-09
Vera Spitzlei
H WS AU MB A