FlexStream™ HG
Humidification Module
Description
The FlexStream™ HG is used in the FlexStream™ Gas
Standards Generating System to add humidity to the trace
concentration gas mixture. However, the technique for adding
the moisture is critical if the integrity of the mixture is to be
maintained. If it is bubbled through water, or even contacts the
liquid surface its concentration may be changed by loss of analyte to the water.
Operation
The FlexStream™ HG has five modes of operation; Offline, Standby, Zero, Span, and Purge, which are
controlled through the microprocessor subsystem on the Base Module. If an installed HG is in the ‘Offline’ mode
the FlexStream™ System ignores its presence. In the ‘Standby’ it ready to use at any time, but but still being
ignored by the System. ‘Zero’ adds humidity to the diluent before the addition of the trace analytes. This mode
is essential in establishing true zero and for standard additions calibrations. The ‘Span’ mode produces humidified,
trace concentration gas standards. The ‘Purge’ mode is used to system clean-up between applications.
To assure accuracy of delivery the HG uses an external humidity sensor installed at the point of use for its control
signal.
MIXTURE
HUMIDIFIED
HUMIDIFIED MIXTURE
OUT
Flow Diagram
SEN
SENSOR
SOR
R
RESERVOIR
ESERVOIR
ANEL
P
PA
FRONT PANEL
CONNECTIONS
CONNECTIONS
R
RECIRCULATING
ECIRCULA
AT
TING
PU
PUMP
MP
AN
EXTERNAL
EXTERNAL SPAN
IN
PI
D
DILUENT
ILUENT IIN
N
(F
(FROM
ROM ""B"UPSTREAM)
B"UPSTREAM)
C
MFC
MFC
ANEL
PANEL
EAR P
REAR
R
ONNECTIONS
CONNECTIONS
C
T
VEN
VENT
PREHEAT
PR
EH
HEA
AT
T
HUMIDIFIER
H
UMIDIFIER
SA
SATURATION
AT
TURA
AT
TION
PR
PRESSURE
ESSURE
CONTROL
NO
Features
• Humidity range ~20% - >90% (max. without secondary dilution)
•
External sensor for real world control
•
Fully automated
•
Compensates for added moisture
•
Standard Flow Range: 0.25 - 5.0slpm
•
No mixture contact with liquid water
•
Membrane separates water from gas – no danger of accidental flooding
•
Elevated saturation temperature
•
Saturation pressure adjustable from ~0psig – 20psig
•
Full control over the useful range
•
Dimensions: 7.5” Width X 13.5” Height X 20” Depth; Weight: Approximately 30lbs.
•
Power Requirements: 115 VAC, 2A; 230 VAC, 1A