CARAT I pro 1.000 Kurz ENG_2014-06-02_01.indd

CARAT I pro
ROUTINE CHECKS AND MAINTENANCE WORK
Ventilator
Quality makes the Difference
Device type 9LV103
What?
When?
By whom?
Before commissioning
Safety check
Weekly
Clean/replace coarse filter,
visual inspection of fine filter
User
Monthly or if heavily
contaminated
Replace fine filter
User
Monthly
Battery quick test
Users/Provider
Every 6 months during
storage
Charge battery to 100 %
Users/Provider
Every 12 months
This document does not replace
the user’s manual. Make sure to
read the user’s manual!
as of device software 1.000
Provider/Service
DEVICE DESCRIPTION
Maintenance 1
Safety check
Provider/Service
Maintenance 2
Provider/Service
After 15000 h blower run Maintenance 5
time or every 5 years
Provider/service
Every 2 years
Brief instructions
1.
Make sure that the battery is fully charged (100 %).
2.
Disconnect the device from the power supply and operate it for 1 hour on battery.
3.
The battery passed the test, if the battery charge is higher than > 10 % after one hour
and the "Low Internal Battery" message has not appeared. Should the battery charge
fall below 10 % and the alarm "Internal battery low" message appear, then the battery
has to be replaced by an authorized service representative.
3
4
5
6
14
11
15
O2
12
8
13
SpO2
DC
9
Com
23
10
©2014 HOFFRICHTER GmbH
All rights reserved.
HOFFRICHTER GmbH
Mettenheimer Straße 12 / 14
19061 Schwerin
Germany
Telephone: +49 385 39925 - 0
Fax:
+49 385 39925 - 25
Email:
[email protected]
www.hoffrichter.de
1
Tube circuit connection - Inspiration
2
Confirm audio alarm pausing for
2 min
Article no. 50000613
CARAT I pro 1.000 Kurz ENG_2014-06-02_01
22
21
20
19
14 Switch device On/Off
15
DC connection
16 SpO2
SpO2 sensor connection
3
Alarm LED
17
SD card slot
4
Power LED
18
Oxygen supply connection
5
Battery LED
19
Oxygen output
6
Start/stop ventilation
20
USB interface (PC connection)
21
Remote alarm/nurse call
RS232 interface (accessories)
7
The HOFFRICHTER GmbH reserves the right to change or replace the brief instructions without prior
notice. Make sure that you are always using the most current version. Please obtain information about the
current version by contacting your provider or by visiting our website at www.hoffrichter.de.
18
17
16
I
0
7
RUNNING A BATTERY QUICK TEST
The battery quick test should be done monthly as follows:
2
1
Multifunctional key MFK
8
Switch to home screen
22
9
Escape
23 Filter cassette
10
 MFK: Unlock/lock keys
11
Control tube connection
12
Oxygen sensor cable connection
13
Pressure measuring tube
connection
MENU STRUCTURE
SYSTEM STRUCTURE
60
A
C
G
B
H
D
Monitoring
Parameter
Alarm Log
System
Statistics
Service
I
O
J
PA
Graphs
Loops
12
1:1,5
f
[hPa]
p
0,570
[bpm]
[%]
SpO2
L
K
15
18
21
0
3
6
9
12
15
18
21
24
27
307
-600
V
[l]
27
30-7
0
24
600
300
V
0
0,100
0,200
0,300
0,400
PCV
0
[l]
22:33
-600
0
3
6
9
12
15
18
21
24
27
30
t
[s]
22:33
Set 1
Set 2
Set 3
Mode
PCV
IPAP
20,0 hPa
PEEP
5,0 hPa
Frequency
12 bpm
Time Inspiration
2,0 s
Ramp
19,8
p
Off
Sigh Function
Off
High Inspiration Volume
[ bpm ]
1:1,5 0,570
I:E
VI
To enable a screen:
1.
[l]
6,840
MV
FiO2
98
SpO2
[%]
Selected screen symbol
 Black
[%]
76
1,00 l
HR
Navigate to the desired screen by
turning the MFK.
[l]
60
3
Minimum Volume
12
f
[ hPa]
[ bpm ]
22:33
Alarm
IMPORTANT OPERATING ACTIONS
"FiO2-Monitoring" setting has been
set to "External" in the system screen.
The oxygen concentration should be
measured with an external FiO2 monitoring device!
Trigger lock "On"
Trigger lock currently active
LOCK/UNLOCK KEYS
DISPLAY BATTERY STATUS
Press the safe key
.
flashes in the toolbar for about 5 s.
Press the MFK during that time.
Touch the battery symbol
SWITCHING ON/OFF THE DEVICE
The device is switched on and off with the
main switch on the back of the device.
Counter ''Audio alarm paused''
No SD card in the device
Press the ON/OFF key
tilation.
PC is connected via the USB port
Error detected
.
04.07.13
22:35
Low Internal Battery
04.07.13
23:16
Low Inspiration Volume
05.07.13
02:24
Low Pressure
05.07.13
05:04
Power Failure
05.07.13
21:55
High Minute Ventilation
06.07.13
02:46
Stenosis
08.07.13
03:24
Low Inspiration Volume
12.07.13
00:10
Disconnection
12.07.13
04:12
Check Expiration Outlet
18.07.13
22:18
Low Pressure
Not selected symbol
 Blue
2.
PCV
22:33
Battery State
Calibrate Tube
FiO2-Monitoring
90 %
charging...
Internal
Calibrate FiO2 Sensor
Alarm Volume
3
Brightness Display
50 %
Brightness LEDs
10 %
Brightness MFK
10 %
Language
English
Pressure Unit
Press the MFK to Continue
hPa
Display Time Insp.
Seconds
PCV
22:33
Statistics
5%
25%
Median
75%
95%
60
to start the ven-
STOP VENTILATION
Press the ON/OFF key
. Select “Yes” by
turning the MFK. Press the MFK.
25
DISPLAY ERROR LIST
Touch the error symbol
Error
Error FiO2-Sensor
Primary Alarm Not Working
f
.
[ bpm ]
20
15
10
0
08 09 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 21 01 02 03 04 05 06
PCV
Jan/14
Feb/14
22:33
Service
Calibrate Pressure Sensor
Calibrate O2 Sensor
Internal battery fully charged
Calibrate Tube
Battery Check
Press the MFK to Continue
Internal battery being charged
(1 bar ≙ 20 % of charge)
O2-Valve Check
Tightness Check
Change PIN
Factory Settings
PCV
Press the MFK to activate the
selected screen.
TIP: The screens can be activated by
touching a screen symbol.
System settings
60
START VENTILATION
FiO2 sensor connected
12
22:33
60
M
SD card is inserted in the device
SpO2 sensor connected
9
[bpm]
HR
Calibrate Flow Sensor
Sp
O2
6
76
Ventilation and alarm parameters
Audio alarm paused
Spontaneous breathing detected
3
PCV
PCV
Oxygen being supplied
0
-300
98
[%]
FiO2
I:E
[l]
MV
60
Keylock activated
600
Flow
[l/min] 300
0
6,840
[l]
VI
NT
P
EXPLANATION OF THE TOOLBAR SYMBOLS
1:45
14
12
19,8
Low-priority alarm
Loops
21
-300
60
Medium-priority alarm
18
0
Measurements
Optional
High-priority alarm
p
[hPa]
[l/min]
6
Monitoring
60
Graphs
24
F
E
Measurements
60
22:33
PCV
Home mode active
Loops
Flow
PCV
N
Clinic mode active
Graphs
28
1
Measurement
60
TIE
Mask
Expiration valve
Tube circuit
Connection tube1
Bacterial filter1
Bacterial filter adapter1
Connector cable1
FiO2 sensor1
FiO2 sensor adapter1
T adapter1
SpO2 sensor1
Control tube
Pressure measuring tube
Humidifier1
Air outlet
Tracheal cannula
A
B
C
D
E
F
G
H
I
J
K
L
M
N
O
P
22:33