Greenspan DO100 Dissolved Oxygen Sensor

DO100
DISSOLVED
OXYGEN SENSOR
PENTAIR ENVIRONMENTAL SYSTEMS
DO100
DISSOLVED OXYGEN SENSOR
DESCRIPTION
The DO100 Dissolved Oxygen
Sensor provides accurate long-term
measurements under a wide range of
environmental conditions.
The patented diffusion cell technology
developed by Greenspan and CSIRO,
provides a simple, effective and robust
means of obtaining continuous in-situ
dissolved oxygen measurements with
minimum maintenance requirements.
A combination of innovative membrane
and sensor technology characterise
the design of the Greenspan Dissolved
Oxygen Sensor. The new generation
chemical sensor provides inherently
greater stability than conventional
polarographic technology. Compared to
conventional membrane designs, the
large surface area of the Greenspan
diffusion membrane substantially
reduces errors caused by surface
contamination. Signal conditioning
electronics are designed with
temperature compensation to provide
true output in ppm of dissolved oxygen
or percent saturation.
DO100
The sensor is ideal for use in a wide
range of monitoring and environmental
projects. A durable corrosion resistant
body and double O ring sealing ensures
suitability for harsh environment
applications. The sensor comes fitted
with a moulded polyurethane cable.
BENEFITS
• Low cost of ownership and minimal
maintenance
The DO100 is easy to install and
provides users with a rugged and
reliable solution to long term
monitoring. For more information
please contact your nearest sales
office.
• Tidal and estuarine studies
FEATURES
• Sensor design requires no moving
parts, ensuring long term reliable
field operation and low power
consumption.
• New generation chemical sensor
provides excellent linearity and
repeatability.
• Sensor output includes
• Ease of use and long term reliability
APPLICATIONS
• Ground water monitoring
• Effluent monitoring
• River and stream measurements
CUSTOMER SUPPORT
Greenspan also offer a range of
customer support functions:
• Field installation
• Data management
• Project integration
• Technical and sales support
• Factory calibration and
re-certification
• Product servicing and repair
supplementary water temperature
reading.
• A response time (T90) of 40 minutes
is achievable by the sensor when
deployed in reasonable flow
47.0
DIMENSIONS
27.0
161.0
330.0
573.0
02
GREENSPAN
55.0
DO100
DISSOLVED OXYGEN SENSOR
DO100 SPECIFICATIONS
Measurement technique
Galvanic cell and patented diffusion rod
Dissolved oxygen range
0–20 ppm or 0–200% saturated
Output
DO 4–20 mA/Temperature 4–20 mA RS232
Dissolved oxygen accuracy
+/– 0.3 ppm (@25°C)
Temperature accuracy
+1°C
Dissolved oxygen linearity
0.2 ppm or ±1% FS
Cable type
Polyurethane sheathed cable, OD 8 mm, Aramid reinforced, moulded entry, bare wire connection
Standard cable lengths
10, 20, 30, 50, 100, 150 m (32, 65, 100, 165, 325, 490 ft)
Non-standard cable lengths
Yes (Extra cable moulding time may be required)
Power supply
9–27 V
Reverse polarity protected
Yes
Surge protected
to 2 kV
Warm-up time to stable reading
1 sec
Current consumption
10–40 mA
Temperature range
0–55°C
Storage temperature
–5°C–60°C
Response time (T90)
40 mins
Depth rating
100 m
Weight
1.66 kg (Stainless Steel)
850 g (Acetal)
Dimensions
Length 573 mm (22.56″ × 1.85″)
OD 47 mm (Stainless Steel or Acetal)
Wetted materials
316 Stainless Steel, Acetal
Silicon Diffusion Membrane
OPTIONAL EXTRAS AND
ACCESSORIES
• Full copper shroud with copper cap
(Part No. 7DR-100A)
• Full acetal shroud with acetal cap
(Part No. 7DR-100D)
• Full acetal shroud with copper cap
(Part No. 7DR-100)
HOW TO ORDER
The following information will be
required:
• Model DO100
• Cable length (m)
• Range (0–20 ppm or
0–200% saturation)
• Shroud option
GREENSPAN
03
Queensland
Phone: +61 1800 805 372
Fax:
+61 1300 658 799
South Australia
Phone: +61 1800 805 372
Fax:
+61 1300 658 799
Victoria
Phone: +61 1800 805 372
Fax:
+61 1300 658 799
Western Australia
Phone: +61 1800 805 372
Fax:
+61 1300 658 799
Greenspan
21 Lawson Crescent
Coffs Harbour NSW 2450
Phone: +61 0 2 6651 9830
Fax:
+61 0 2 6651 9831
70 Cleaver Terrace
Belmont WA 6104
Phone: +61 0 8 9477 1188
Fax:
+61 0 8 9479 6727
EUROPE
United Kingdom
Pentair Environmental Systems Ltd
Unit 3B Beechwood
Chineham Business Park
Basingstoke, Hampshire
RG24 8WA
United Kingdom
Phone: +44 1256 817 800
Fax:
+44 1256 332 760
Germany
Pentair Umwelttechnik GmbH
Im Petersfeld 6
D-65624 Altendiez
Germany
Phone: +49 6432 95299-0
Fax:
+49 6432 95299-24
Italy
Mecair s.r.l
Via per Cinisello 97
20834 Nova Milanese
Monza e Brianza, Italy
Phone: +39 0362 3751
Fax:
+39 0362 367 279
ASIA
China
21/F, Cloud 9 Plaza
1118 West Yan An Road
Shanghai 200052
China
Phone: +86 21 5239 8810
Fax:
+86 21 3211 4582
Malaysia
73-M Jalan Mega Mendung
Kompleks Bandar, OUG,
58200 Kuala Lumpur
Malaysia
Phone: +60 3 7987 6839
Fax:
+60 3 7987 7839
Greenspan
Blk 13 Kaki Bukit Road 1
Eunos Technolink #03-03
Singapore 415928
Singapore
Phone: +65 6748 0140
Fax:
+65 6748 1948
USA
Corporate Office
Goyen Valve LLC
1195 Airport Rd
Lakewood, NJ 08701
732-364-7800
22 Palmerin Street
Warwick QLD 4370
Phone: +61 7 4660 1888
Fax:
+61 7 4660 1800
CEMS & Process Gas Monitoring
Sales, Technical Support & Proposals
Gary Cacciatore
714-538-7212
[email protected]
240 Lavarack Avenue
Eagle Farm QLD 4009
Phone: +61 7 3866 7829
Fax:
+61 7 3260 1916
Service & Support
Don Ribbens
402-680-6669
[email protected]
CLEANAIRSYSTEMS.COM
© 2013 Pentair Environmental Systems reserves the right to change product designs and specifications without notice.
DO100 Rev03 04/13
AUSTRALIA
New South Wales – Head Office
268–292 Milperra Rd
Milperra NSW 2214
Phone: +61 1800 805 372
Fax:
+61 1300 658 799