USING MASS SPECTRAL DETECTION WITH USP METHODS Authors: Thomas E. Wheat, Daniel S. Root, Aparna Chavali, Patricia McConville Affiliations: Waters Corporation, Milford MA, USA INTRODUCTION The analytical methods that are included in USP monographs are preferred for the development and quality control of drug substances and products. While the chromatographic methods are typically robust and often include the common related substances, they can be difficult to validate where confirmation of peak identity is required. Further, identification of unexpected components can prove difficult. These complications arise because the specified mobile phases often contain high concentrations of buffer salts. These conditions compromise ionization for mass spectral detection and can complicate collection of fractions for use in complementary techniques. Substantial effort has been expended in many laboratories in converting these USP methods to mobile phase modifiers more suitable for MS. This challenging redevelopment of methods has seldom proven efficient or satisfactory. Further, that process would require substantial revalidation of the chromatographic method. As a more practical alternative, a systematic protocol has been developed for using MS detection directly with these methods and for guiding the transition to generally useful separations eliminating non-volatile buffers. METHOD System ACQUITY® UPLC System with 2D Technology Alpha Pump: ACQUITY UPLC H-Class Quaternary Solvent Manager QSM Beta Pump: ACQUITY UPLC I-Class Binary Solvent Manager BSM Dilution Pump: ACQUITY Isocratic Solvent Manager ISM Sample Manager: ACQUITY UPLC I-Class Sample Manager – Flow-ThroughNeedle Column Manager (CM-A) configured for two Active Pre-Heaters, additional tubing and 2D valve kit UV Detection: ACQUITY UPLC Photodiode Array (PDA) Detector 220 nm Mass Detection: ACQUITY QDa Detector MassLynx 4.1 SCN 888 with Waters pump control Figure 1. Plumbing diagram for peak transfer from a USP HPLC method. Red path indicates flow through the HPLC column. The blue indicates flow through the UPLC column. MassLynx system software controls all module and valve events. The ISM flow combines with the flow from the HPLC column to dilute organic concentration for retention of the transferred peak. Figure 4. ESI+ SIR confirmation of the identity of Irbesartan API at m/z 429.2 from peak transfer. The baseline upset at 20.5 min. reflects the flow change with trapping. Figure 2. Irbesartan USP UV profile. Unidentified peak is most likely related compound A, but confirmation is needed. This peak will be transferred to the mass detector. Figure 5. The ESI+ SIR at m/z 447.1 confirms the transferred, unidentified peak as related compound A. Conditions USP method HPLC column: Mobile phase: Solution: XSelect HSS T3 4.6mm x 250mm 5 μm (L1) Buffer Solution and Acetonitrile (60:40), Buffer 0.55% Phosphoric Acid in Water adjusted to pH 3.2 with Triethylamine , Isocratic Flow rate: 1.0 mL/min UPLC elution method UPLC column: BEH C18, 2.1 x 50 mm, 1.7µm Mobile phases: Water, Acetonitrile Dilution flow: 0.1% formic acid in water, 0.2 mL/min Gradient: 0 to 95% acetonitrile, 3 minutes, 0.5 mL/min Mass detection ESI+ SIR Irbesartan m/z 429.2 , and Irbesartan related compound A (RCA) m/z 447, Full scan m/z 100—500 CONCLUSION PROTOCOL USP method is run with UV detection—QSM delivering flow When peak of interest begins, flow is diverted to the UPLC Column flow from the Beta pump is delivered to the HPLC column flow from the ISM dilutes the flow prior to the UPLC column flow from the UPLC column is sent to waste the peak is trapped at the head of the UPLC column The USP mobile phase is eluted to waste and replaced With the flow sent to the mass detector, the compound of interest is eluted from the UPLC column into the QDa mass detector The QDa detector monitors the eluate using SIR for expected compounds, with full scan to detect additional species The fully-automated multi-valve, multi-pump ACQUITY® UPLC System can be configured to combine analytical separations At-column dilution ensures good retention and peak shape of analytes transferred with this protocol The described configuration and protocol can be used for mass spectral detection of analytes separated with high ionic strength, non-volatile mobile phases without redeveloping and revalidating the method This method can be applied to the identification of minor or unidentified components in pharmaceutical products Figure 3. The benefit of at-column dilution (ACD) for the transferred peak of related compound A. Without ACD the peak is retained but very distorted. TO DOWNLOAD A COPY OF THIS POSTER, VISIT WWW.WATERS.COM/POSTERS ©2014 Waters Corporation
© Copyright 2024 ExpyDoc